- Indira Narayanan, Marvesh Mendhi, Pooja Bansil, and Patricia S Coffey.
- Department of Pediatrics/Neonatology, Georgetown University Medical Center, Washington, D.C.
- Respir Care. 2017 Nov 1; 62 (11): 1428-1436.
BackgroundThe study assessed the impact of simulated ventilation techniques using upright and conventional self-inflating neonatal resuscitators on delivered tidal volume (VT) and pressure.MethodsWe analyzed videos of participants ventilating a manikin using an upright (upright, n = 33) and a conventional resuscitator (conventional, n = 32) under normal and low lung compliance. Mask hold, number of fingers squeezing the bag, and degree of bag squeeze were compared with VT and peak inspiratory pressure (PIP).ResultsVT and PIP values were higher when using the upright resuscitator than when using the conventional resuscitator. With low compliance, differences in VT were insignificant except with the use of the OK/C hold, (upright, 29.6 ± 4.0 mL, vs conventional, 24.8 ± 6.0 mL, P = .02). PIP was significantly higher when using the upright resuscitator with the OK hold (upright, 36.3 ± 4.4 mL, vs conventional, 30.3 ± 6.6 mL, P = .009) and when the bag was squeezed by more than half (upright, 33.8 ± 16.3 mL, vs conventional, 29.3 ± 9.5 mL, P = .046). With normal compliance, VT was high with both resuscitators, being significantly higher when using the upright resuscitator with the OK hold (upright, 64.3 ± 9.5 mL, vs conventional, 45.8 ± 9.4 mL; P < .001), and when the bag was squeezed using more than 2 fingers (upright, 58.0 ± 17.2 mL, vs conventional, 45.7 ± 12.6 mL, P = .01) and by more than half (upright, 58.7 ± 16.6 mL, vs conventional, 45.8 ± 12.2 mL, P = .004). PIP, too, was significantly higher when using the upright resuscitator with the OK hold (upright, 29.3 ± 3.5 mL, vs conventional, 21.5 ± 4.0 mL, P = <.001) and when the bag was squeezed using more than 2 fingers (upright, 27.2 ± 7.0 mL, vs conventional, 21.6 ± 5.7 mL, P = .005), and by more than half (upright, 27.6 ± 6.6 mL, vs conventional, 21.7 ± 5.4 mL, P = .001).ConclusionsImproved mask design, larger bag volume, and upright orientation of the upright resuscitator likely contributed to higher VT and PIP. However, high VT was observed with both resuscitators, possibly due to excessive squeezing of the bag, especially during low compliance. Thus, the design of the resuscitator and manner in which the device is utilized can both significantly influence the VT and PIP attained.Copyright © 2017 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.