    Comparative Study

    Signify ER Drug Screen Test evaluation: comparison to Triage Drug of Abuse Panel plus tricyclic antidepressants.

    • Jane Ellen Phillips, Stuart Bogema, Paul Fu, Wieslaw Furmaga, Alan H B Wu, Vlasta Zic, and Catherine Hammett-Stabler.
    • University of North Carolina Hospitals, McLendon Clinical Labs, Department of Pathology and Laboratory Medicine, 101 Manning Drive, Chapel Hill, NC 27514, USA. jane.phillips@roche.com
    • Clin. Chim. Acta. 2003 Feb 1; 328 (1-2): 31-8.

    AbstractSignify ER Drug Screen Test (Signify ER) and Triage Drug of Abuse Panel plus TCA (Triage DOA Panel) rapid drug screening devices were compared at four laboratories. Both assay systems are point of care immunoassays, measuring phencyclidine, barbiturates, amphetamine, cocaine metabolite, methamphetamine, tricyclic antidepressants, opiates, marijuana metabolite, and benzodiazepines in human urine. The performance of these two assay systems, including a cutoff verification and cross-reactivity using spiked urine specimens and accuracy using clinical urine samples, was investigated. The cutoff verification study showed that the Signify ER had 95.4% precision for all drugs tested at concentrations of 50%, 75%, 125%, 150%, and 200% of cutoffs compared to 90% precision obtained with Triage DOA Panel. Accuracy studies testing 53 negative urine samples demonstrated that both Signify ER and Triage DOA Panel have 100% specificity. Testing of 693 positive urine samples demonstrated that Signify ER and Triage DOA Panel have sensitivities of 99.8% and 99.3%, respectively, with an accuracy of 99.9% and 99.6%. A total of 527 compounds were tested for the cross-reactivity study. Eighty-seven structurally related drugs and metabolites were found to cross-react with at least one of the nine tests of the Signify ER. Four hundred forty structurally unrelated compounds that can be found in human urine were shown not to cross-react with the Signify ER. In terms of operating characteristics, the Signify ER device is simpler since only a single pipetting step is required, and reaction completed within 8 min.Copyright 2002 Published by Elsevier Science B.V.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.