    Comparative Study Clinical Trial Controlled Clinical Trial

    Defibrillation efficacy of commercially available biphasic impulses in humans. Importance of negative-phase peak voltage.

    • G Tomassoni, K Newby, S Deshpande, K Axtell, J Sra, M Akhtar, and A Natale.
    • Electrophysiology Laboratory, Duke University, VA Medical Center, Durham, NC 27705, USA.
    • Circulation. 1997 Apr 1; 95 (7): 1822-6.

    BackgroundRecent studies have shown that specifically shaped biphasic waveforms can lower energy requirements for ventricular defibrillation. We prospectively compared the defibrillation efficacy of three different biphasic wave shapes incorporated in three commercially available implantable defibrillators. The results led to the development of a second protocol in which the importance of negative-phase peak voltage and duration was investigated.Methods And ResultsDefibrillation threshold (DFT) testing using different biphasic waveforms was performed randomly on 42 patients undergoing implantation of a cardioverter-defibrillator for ventricular arrhythmias. In 23 patients (group 1), 3 waveforms were tested: a CPI waveform with 60% positive-phase (P1) tilt and 50% negative-phase (P2) tilt, a Medtronic waveform with 65% fixed tilt in both P1 and P2, and a Ventritex waveform with 60% P1 tilt and a P2 leading edge voltage equal to half of the P1 trailing edge voltage. In 19 patients (group 2), 3 biphasic waveforms with equal P1 tilt at 65% but shorter P2 duration or smaller P2 peak voltage were tested. The Endotak C 60 series lead system (CPI) was used in 11 patients in group 1 and 10 patients in group 2. A Transvene lead system (Medtronics) was used in the remaining patients. Stored energy required for defibrillation was significantly lower with the CPI waveform compared with the Ventritex waveform. In group 2, energy requirements were significantly increased for the waveform with a smaller P2 peak voltage, whereas a short P2 duration did not influence defibrillation success.ConclusionsOur results suggest that specifically shaped biphasic waveforms delivered from commercially available devices can affect energy requirements for defibrillation. More importantly, the amplitude of the P2 peak voltage may be a more critical determinant than the P2 duration for defibrillation success of biphasic waveforms in humans.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…