J Clin Monit Comput · Aug 2008
Non-invasive automated measurement of cardiac output during stable cardiac surgery using a fully integrated differential CO(2) Fick method.
- Philip J Peyton, Daniel Thompson, and Paul Junor.
- Department of Anaesthesia, Austin Hospital, University of Melbourne, Melborne, Australia. phil.peyton@austin.org.au
- J Clin Monit Comput. 2008 Aug 1; 22 (4): 285-92.
ObjectivesTo re-evaluate the accuracy and precision of a non-invasive method for measurement of cardiac output based on the differential CO(2) Fick approach using an automated change in respiratory rate delivered by a ventilator under control by a prototype measurement system.MethodsTwenty-four patients during coronary artery bypass surgery, pre- and postcardiopulmonary bypass were recruited. After routine cannulation including pulmonary artery catheter, relaxant general anesthesia was induced. After hemodynamic and ventilatory stability were achieved, simultaneous paired measurements were made by the differential Fick method and by bolus thermodilution. Measurements were generated by inducing a change in respiratory rate by the ventilator under computer control. In Group 1, this involved an increase in respiratory rate from 8 to 12 breaths/min. In Group 2, this involved a decrease from 12 to 6 breaths/min.ResultsNineteen measurements were made in each Group, 12 pre-CPB and 7 post-CPB. In Group 1 mean bias was -0.06 l/min, with a precision of agreement of 0.87 l/min, r = 0.91. In Group 2 (excluding one outlier) mean bias was -0.07 l/min, with a precision of 1.12 l/min, r = 0.71.ConclusionsAcceptable agreement with thermo- dilution during surgery was found, particularly where the ventilatory change involved an increase in respiratory rate from a lower baseline. This approach has potential to be readily integrated into modern anesthesia delivery platforms, allowing routine non-invasive cardiac output measurement.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.