• J Spinal Disord Tech · Aug 2008

    Correlation of ProDisc-C failure strength with cervical bone mineral content and endplate strength.

    • Xingkai Zhang, Nathaniel R Ordway, Rong Tan, Byeong Cheol Rim, and Amir H Fayyazi.
    • Shanghai Jiaotong University Medical College Ruijin Hospital, Shanghai Institute of Orthopedics and Traumatology, Shanghai, China.
    • J Spinal Disord Tech. 2008 Aug 1; 21 (6): 400-5.

    Study DesignThis study was designed to evaluate the effectiveness of preoperative computed tomography (CT) scan in predicting endplate strength.ObjectiveTo demonstrate a correlation between the cervical trabecular bone density and the failure strength of the endplate.Summary Of Background DataCervical total disc arthroplasty devices have to transmit the force to the endplate for the remainder of the patients' life. One potential complication at this interface is endplate fracture and implant subsidence, which usually occurs early postoperatively and may be related to weakness of the boney endplate.MethodsSix fresh human cadaver cervical spines were harvested and scanned for the determination of trabecular density using a peripheral quantitative CT (QCT) scanner. The specimens were then disarticulated and the inferior endplate of each vertebral segment was biomechanically tested using a 2-mm indentation probe to determine average endplate strength. The superior endplate of each vertebral body was then fitted with the appropriately sized ProDisc-C endplate and biomechanically tested until failure. Regression analyses were used to compare the interface failure stress of the implant with the bone mineral content and the average endplate stress as measured with the indentation probe.ResultsThe average bone mineral content of the specimens was 322+/-57 mg/cm. The average endplate strength and stress measured by the indentation test was 176+/-129 N and 56+/-34 N/mm, respectively. The average ProDisc-C/endplate failure load and failure stress were 1875+/-1023 N and 10.2+/-4.1 N/mm, respectively. There was a direct correlation between the ProDisc-C/endplate failure stress and the bone mineral content measured by peripheral QCT (R=0.48, P<0.01). There was also a significant correlation between ProDisc-C/endplate failure stress and the endplate indentation stress.ConclusionsThis study demonstrates the utility of a preoperative QCT scan in predicting the failure stress of the cervical endplate before total disc replacement. This information may potentially decrease early complications of device subsidence or endplate fracture.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…