-
- Frédéric Duprez, Jean Bernard Michotte, Gregory Cuvelier, Alexandre Legrand, Sharam Mashayekhi, and Gregory Reychler.
- Service des Soins Intensifs, Hôpital Epicura, Site Hornu, Belgium; Institut de Recherche Expérimentale et Clinique (IREC), Pôle de Pneumologie, ORL & Dermatologie, Service de Pneumologie, Université Catholique de Louvain, Brussels, Belgium; and Laboratoire de l'Effort et du Mouvement, Condorcet, Tournai, Belgium.
- Respir Care. 2018 Mar 1; 63 (3): 332-338.
BackgroundOxygen cylinders are widely used both in hospital and prehospital care. Excessive or inappropriate FIO2 may be critical for patients with hypercapnia or hypoxia. Moreover, over-oxygenation could be deleterious in ischemic disorders. Supplemental oxygen from oxygen cylinder should therefore be delivered accurately. The aim of this study was to assess the accuracy of oxygen flows for oxygen cylinder in hospital and prehospital care.MethodsA prospective trial was conducted to evaluate accuracy of delivered oxygen flows (2, 4, 6, 9 and 12 L/min) for different oxygen cylinder ready for use in different hospital departments. Delivered flows were analyzed randomly using a calibrated thermal mass flow meter. Two types of oxygen cylinder were evaluated: 78 oxygen cylinder with a single-stage regulator and 70 oxygen cylinder with a dual-stage regulator. Delivered flows were compared to the required oxygen flow. The residual pressure value for each oxygen cylinder was considered. A coefficient of variation was calculated to compare the variability of the delivered flow between the two types of oxygen cylinder.ResultsThe median values of delivered flows were all ≥ 100% of the required flow for single stage (range 100-109%) and < 100% of required flow for dual stage (range 95-97%). The median values of the delivered flow differed between single and dual stage. It was found that single stage is significantly higher than dual stage (P = .01). At low flow, the dispersion of the measures for single stage was higher than with a high oxygen flow. Delivered flow differences were also found between low and high residual pressures, but only with single stage (P = .02). The residual pressure for both oxygen cylinders (no. = 148) ranged from 73 to 2,900 pounds per square inch, and no significant difference was observed between the 2 types (P = .86). The calculated coefficient of variation ranged from 7% (±1%) for dual stage to 8% (±2%) for single stage.ConclusionsThis study shows good accuracy of oxygen flow delivered via oxygen cylinders. This accuracy was higher with dual stage. Single stage was also accurate, however, at low flow this accuracy is slightly less. Moreover, with single stage, when residual pressure decreases, the median value of delivered flow decreased.Copyright © 2018 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.