- Meera Chand, Theresa Lamagni, Katharina Kranzer, Jessica Hedge, Ginny Moore, Simon Parks, Samuel Collins, Carlos Del Ojo Elias, Nada Ahmed, Tim Brown, E Grace Smith, Peter Hoffman, Peter Kirwan, Brendan Mason, Alison Smith-Palmer, Philip Veal, Maeve K Lalor, Allan Bennett, James Walker, Alicia Yeap, Antonio Isidro Carrion Martin, Gayle Dolan, Sonia Bhatt, Andrew Skingsley, André Charlett, David Pearce, Katherine Russell, Simon Kendall, Andrew A Klein, Stephen Robins, Silke Schelenz, William Newsholme, Stephanie Thomas, Tim Collyns, Eleri Davies, Jim McMenamin, Lorraine Doherty, Tim E A Peto, Derrick Crook, Maria Zambon, and Nick Phin.
- National Infection Service, Public Health England.
- Clin. Infect. Dis. 2017 Feb 1; 64 (3): 335-342.
BackgroundAn urgent UK investigation was launched to assess risk of invasive Mycobacterium chimaera infection in cardiothoracic surgery and a possible association with cardiopulmonary bypass heater-cooler units following alerts in Switzerland and The Netherlands.MethodsParallel investigations were pursued: (1) identification of cardiopulmonary bypass-associated M. chimaera infection through national laboratory and hospital admissions data linkage; (2) cohort study to assess patient risk; (3) microbiological and aerobiological investigations of heater-coolers in situ and under controlled laboratory conditions; and (4) whole-genome sequencing of clinical and environmental isolates.ResultsEighteen probable cases of cardiopulmonary bypass-associated M. chimaera infection were identified; all except one occurred in adults. Patients had undergone valve replacement in 11 hospitals between 2007 and 2015, a median of 19 months prior to onset (range, 3 months to 5 years). Risk to patients increased after 2010 from <0.2 to 1.65 per 10000 person-years in 2013, a 9-fold rise for infections within 2 years of surgery (rate ratio, 9.08 [95% CI, 1.81-87.76]). Endocarditis was the most common presentation (n = 11). To date, 9 patients have died. Investigations identified aerosol release through breaches in heater-cooler tanks. Mycobacterium chimaera and other pathogens were recovered from water and air samples. Phylogenetic analysis found close clustering of strains from probable cases.ConclusionsWe identified low but escalating risk of severe M. chimaera infection associated with heater-coolers with cases in a quarter of cardiothoracic centers. Our investigations strengthen etiological evidence for the role of heater-coolers in transmission and raise the possibility of an ongoing, international point-source outbreak. Active management of heater-coolers and heightened clinical awareness are imperative given the consequences of infection.© Crown copyright 2016.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.