• Spine · Jul 2018

    Feasibility and Accuracy of Thoracolumbar Minimally Invasive Pedicle Screw Placement With Augmented Reality Navigation Technology.

    • Adrian Elmi-Terander, Rami Nachabe, Halldor Skulason, Kyrre Pedersen, Michael Söderman, John Racadio, Drazenko Babic, Paul Gerdhem, and Erik Edström.
    • Department of Neuroradiology, Karolinska Hospital, Stockholm, Sweden.
    • Spine. 2018 Jul 15; 43 (14): 1018-1023.

    Study DesignCadaveric laboratory study.ObjectiveTo assess the feasibility and accuracy of minimally invasive thoracolumbar pedicle screw placement using augmented reality (AR) surgical navigation.Summary Of Background DataMinimally invasive spine (MIS) surgery has increasingly become the method of choice for a wide variety of spine pathologies. Navigation technology based on AR has been shown to be feasible, accurate, and safe in open procedures. AR technology may also be used for MIS surgery.MethodsThe AR surgical navigation was installed in a hybrid operating room (OR). The hybrid OR includes a surgical table, a motorized flat detector C-arm with intraoperative 2D/3D imaging capabilities, integrated optical cameras for AR navigation, and patient motion tracking using optical markers on the skin. Navigation and screw placement was without any x-ray guidance. Two neurosurgeons placed 66 Jamshidi needles (two cadavers) and 18 cannulated pedicle screws (one cadaver) in the thoracolumbar spine. Technical accuracy was evaluated by measuring the distance between the tip of the actual needle position and the corresponding planned path as well as the angles between the needle and the desired path. Time needed for navigation along the virtual planned path was measured. An independent reviewer assessed the postoperative scans for the pedicle screws' clinical accuracy.ResultsNavigation time per insertion was 90 ± 53 seconds with an accuracy of 2.2 ± 1.3 mm. Accuracy was not dependent on operator. There was no correlation between navigation time and accuracy. The mean error angle between the Jamshidi needles and planned paths was 0.9° ± 0.8°. No screw was misplaced outside the pedicle. Two screws breached 2 to 4 mm yielding an overall accuracy of 89% (16/18).ConclusionMIS screw placement directed by AR with intraoperative 3D imaging in a hybrid OR is accurate and efficient, without any fluoroscopy or x-ray imaging during the procedure.Level Of EvidenceN/A.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…