• Resuscitation · Mar 2018

    Multicenter Study Comparative Study

    Efficacy of different cooling technologies for therapeutic temperature management: A prospective intervention study.

    • Petra Sonder, Gladys N Janssens, Albertus Beishuizen, Connie L Henry, Jon C Rittenberger, Clifton W Callaway, Cameron Dezfulian, and Kees H Polderman.
    • Department of Critical Care Medicine, University of Pittsburgh Medical Center, 200 Lothrop St, Pittsburgh, PA 15213, United States.
    • Resuscitation. 2018 Mar 1; 124: 14-20.

    BackgroundMild therapeutic hypothermia (32-36 °C) is associated with improved outcomes in patients with brain injury after cardiac arrest (CA). Various devices are available to induce and maintain hypothermia, but few studies have compared the performance of these devices. We performed a prospective study to compare four frequently used cooling systems in inducing and maintaining hypothermia followed by controlled rewarming.MethodsWe performed a prospective multi-centered study in ten ICU's in three hospitals within the UPMC health system. Four different cooling technologies (seven cooling methods in total) were studied: two external water-circulating cooling blankets (Meditherm® and Blanketrol®), gel-coated adhesive cooling pads (Arctic Sun®), and endovascular cooling catheters with balloons circulating ice-cold saline (Thermogard®). For the latter system we studied three different types of catheter with two, three or four water-circulating balloons, respectively. In contrast to previous studies, we not only studied the cooling rate (i.e., time to target temperature) in the induction phase, but also the percentage of the time during the maintenance phase that temperature was on target ±0.5 °C, and the efficacy of devices to control rewarming. We believe that these are more important indicators of device performance than induction speed alone.Results129 consecutive patients admitted after CA and treated with hypothermia were screened, and 120 were enrolled in the study. Two researchers dedicated fulltime to this study monitored TH treatment in all patients, including antishivering measures, additional cooling measures used (e.g. icepacks and cold fluid infusion), and all other issues related to temperature management. Baseline characteristics were similar for all groups. Cooling rates were 2.06 ± 1.12 °C/h for endovascular cooling, 1.49 ± 0.82 for Arctic sun, 0.61 ± 0.36 for Meditherm and 1.22 ± 1.12 for Blanketrol. Time within target range ±0.5 °C was 97.3 ± 6.0% for Thermogard, 81.8 ± 25.2% for Arctic Sun, 57.4 ± 29.3% for Meditherm, and 64.5 ± 20.1% for Blanketrol. The following differences were significant: Thermogard vs. Meditherm (p < 0.01), Thermogard vs. Blanketrol (p < 0.01), and Arctic Sun vs. Meditherm (p < 0.02). No major complications occurred with any device.ConclusionsEndovascular cooling and gel-adhesive pads provide more rapid hypothermia induction and more effective temperature maintenance compared to water-circulating cooling blankets. This applied to induction speed, but (more importantly) also to time within target range during maintenance.Copyright © 2017. Published by Elsevier B.V.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…