Blood pressure monitoring · Aug 2017
Multicenter Study Comparative StudyClinical validation of a new wrist continuous noninvasive hemodynamic monitoring system in comparison with invasive radial artery measurement.
- Chen Li, Chanjuan Li, Wei Zhang, Ling Wang, Zhidong Wang, Yongji Wang, and Jielai Xia.
- Department of Health Statistics, School of Preventive Medicine, the Fourth Military Medical University, Xi'an, Shaanxi, People's Republic of China.
- Blood Press Monit. 2017 Aug 1; 22 (4): 238-243.
BackgroundAn ideal arterial pressure monitoring system for perioperative care is required to be accurate, noninvasive, continuous, and risk free. Although several continuous noninvasive arterial pressures (CNAP) are determined on the finger, a new wrist CNAP monitoring system was developed. This prospective study was designed as a randomized-controlled trial to assess its validity in comparison with invasive arterial pressure (IAP) monitoring.Patients And MethodsSixty patients undergoing elective surgery under general anesthesia were enrolled. One-side arm in each patient was selected randomly to assess the new monitoring system placed on the wrist; an arterial catheter for IAP was inserted at the radial artery in the contralateral arm. The Bland-Altman method for repeated measurements was used to assess agreement between measurement methods by the levels of agreement according to the American Association for the Advancement of Medical Instrumentation standards.ResultsA total of 6600 valid pressure readings were obtained from 60 patients, including 3000 beat data and 3600 pulse wave data, respectively. The mean differences in wrist CNAP versus IAP for beat data were as follows: systolic arterial pressure (SAP): -2.09±5.39 mmHg; mean arterial pressure (MAP): 0.27±3.64 mmHg; diastolic arterial pressure (DAP): and 2.63±6.44 mmHg. For pulse wave data, the mean difference (SD) values were as follows: SAP: -2.06±6.51 mmHg; MAP: 0.50±4.36 mmHg; and DAP: 3.06±6.81 mmHg. Percentage errors were less than 28.3%. Levels of agreements were detected to be -12.65-8.47 for SAP, -6.86-7.40 for MAP, and -9.99-15.26 for DAP.ConclusionThe new wrist CNAP monitoring system showed an acceptable agreement and interchangeability with the IAP.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.