    Engineering study comparing injection force and dose accuracy between two prefilled insulin injection pens.

    • Debra A Ignaut, Michael R Opincar, Paula E Clark, Melanie K Palaisa, and Sheila M Lenox.
    • Eli Lilly and Company, Indianapolis, IN 46285, USA. dignaut@lilly.com
    • Curr Med Res Opin. 2009 Dec 1; 25 (12): 2829-33.

    ObjectiveThis study compared injection force (measured by glide force [GF] and glide force variability [GFV]) and dosing accuracy of the Humalog KwikPen * (prefilled insulin lispro [Humalog dagger] pen, Eli Lilly and Company, Indianapolis, IN) and the Next Generation FlexPen double dagger (prefilled insulin aspart [NovoRapid section sign] pen, Novo Nordisk A/S, Bagsvaerd, Denmark). * Humalog KwikPen is a registered trademark of Eli Lilly and Company, Indianapolis, IN, USA. dagger Humalog is a registered trademark of Eli Lilly and Company, Indianapolis, IN, USA. double dagger FlexPen is a registered trademark of Novo Nordisk A/S, Bagsvaerd, Denmark. section sign NovoRapid is a registered trademark of Novo Nordisk A/S, Bagsvaerd, Denmark.Research Design And MethodsA total of 100 prefilled insulin pens (50 insulin lispro pens, 50 insulin aspart pens) were tested using two dose sizes (30 U and 60 U). In all, 50 devices (25 of each type) were tested at 10 U/s dosing speed and 50 were tested at 6.6 U/s. Devices were used per manufacturer instructions. Dose accuracy (represented as absolute dose error %), maximum and average GF, and GFV data were automatically collected by the test system for all datasets (dose size/dosing speed/device type). The test system controlled for potential dosing errors.ResultsThe insulin lispro pen demonstrated a significantly lower median maximum GF at both dosing speeds: (2.83 vs. 3.92 lbs [30 U] and 3.00 vs. 4.14 lbs [60 U]) at 10 U/s; (1.85 vs. 2.93 lbs [30 U] and 2.14 vs. 3.02 lbs [60 U]) at 6.6 U/s, all p < 0.0001. For all datasets, the median GFV was significantly lower for the insulin lispro pen, p < 0.0001. Median dose error was comparable between device types when tested at 10 U/s dosing speed; however, at 6.6 U/s, the median dose error was significantly lower for insulin lispro pen compared to insulin aspart pen (0.47 vs. 0.67% [30 U] and 0.50 vs. 0.78% [60 U], both p < 0.05).ConclusionsThe insulin lispro pen had significantly lower median GF and GFV compared with insulin aspart pen when tested at two dose sizes and two dosing speeds. Median dose error was similar between the device types at the 10 U/s dosing speed, but median dose error was significantly lower for the insulin lispro pen at the 6.6 U/s dosing speed. A limitation of this study was that it was executed as an open label study.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.