• Health devices · Sep 2011

    Medrad Continuum MR infusion system components must remain secured in MR environment.

    • Health Devices. 2011 Sep 1; 40 (9): 312-3.

    AbstractThe infusion pump component of the Medrad Continuum MR infusion system contains ferromagnetic materials and may be pulled into the MR system's bore if not adequately secured. Users need to be aware that neither the pump nor any other components should be removed from the system within the scan room. Additionally, before bringing the system into the scan room, users should verify that the system is free of damage and that the pump and its mounting bracket are both firmly secured.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.