Anesthesia and analgesia · Aug 2019
Comparative StudyHalving the Volume of AnaConDa: Evaluation of a New Small-Volume Anesthetic Reflector in a Test Lung Model.
- Hagen Bomberg, Franziska Meiser, Philipp Daume, Martin Bellgardt, Thomas Volk, Daniel I Sessler, Heinrich V Groesdonk, and Andreas Meiser.
- From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Medical Centre, University of Saarland, Homburg/Saar, Germany.
- Anesth. Analg. 2019 Aug 1; 129 (2): 371-379.
BackgroundVolatile anesthetics are increasingly used for sedation in intensive care units. The most common administration system is AnaConDa-100 mL (ACD-100; Sedana Medical, Uppsala, Sweden), which reflects volatile anesthetics in open ventilation circuits. AnaConDa-50 mL (ACD-50) is a new device with half the volumetric dead space. Carbon dioxide (CO2) can be retained with both devices. We therefore compared the CO2 elimination and isoflurane reflection efficiency of both devices.MethodsA test lung constantly insufflated with CO2 was ventilated with a tidal volume of 500 mL at 10 breaths/min. End-tidal CO2 (EtCO2) partial pressure was measured using 3 different devices: a heat-and-moisture exchanger (HME, 35 mL), ACD-100, and ACD-50 under 4 different experimental conditions: ambient temperature pressure (ATP), body temperature pressure saturated (BTPS) conditions, BTPS with 0.4 Vol% isoflurane (ISO-0.4), and BTPS with 1.2 Vol% isoflurane. Fifty breaths were recorded at 3 time points (n = 150) for each device and each condition. To determine device dead space, we adjusted the tidal volume to maintain normocapnia (n = 3), for each device. Thereafter, we determined reflection efficiency by measuring isoflurane concentrations at infusion rates varying from 0.5 to 20 mL/h (n = 3), for each device.ResultsEtCO2 was consistently greater with ACD-100 than with ACD-50 and HME (ISO-0.4, mean ± standard deviations: ACD-100, 52.4 ± 0.8; ACD-50, 44.4 ± 0.8; HME, 40.1 ± 0.4 mm Hg; differences of means of EtCO2 [respective 95% confidence intervals]: ACD-100 - ACD-50, 8.0 [7.9-8.1] mm Hg, P < .001; ACD-100 - HME, 12.3 [12.2-12.4] mm Hg, P < .001; ACD-50 - HME, 4.3 [4.2-4.3] mm Hg, P < .001). It was greatest under ATP, less under BTPS, and least with ISO-0.4 and BTPS with 1.2 Vol% isoflurane. In addition to the 100 or 50 mL "volumetric dead space" of each AnaConDa, "reflective dead space" was 40 mL with ACD-100 and 25 mL with ACD-50 when using isoflurane. Isoflurane reflection was highest under ATP. Under BTPS with CO2 insufflation and isoflurane concentrations around 0.4 Vol%, reflection efficiency was 93% with ACD-100 and 80% with ACD-50.ConclusionsIsoflurane reflection remained sufficient with the ACD-50 at clinical anesthetic concentrations, while CO2 elimination was improved. The ACD-50 should be practical for tidal volumes as low as 200 mL, allowing lung-protective ventilation even in small patients.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.