Catheter Cardiovasc Interv · Sep 2013
Case ReportsEarly percutaneous valve failure within bioprosthetic tricuspid tissue valve replacements.
- James Bentham, Shakeel Qureshi, Andreas Eicken, John Gibbs, George Ballard, and John Thomson.
- Yorkshire Heart Centre, Leeds General Infirmary, Leeds, United Kingdom.
- Catheter Cardiovasc Interv. 2013 Sep 1; 82 (3): 428-35.
ObjectiveTo describe the cases of percutaneous valve failure following implantation of the Medtronic Melody™ and the Edwards Sapien™ valve in the tricuspid position.BackgroundThere have been recent reports of the use of percutaneously delivered valves to manage failing bioprosthetic tricuspid valves. Total patient numbers remain small and follow-up times are limited.MethodsWe describe four procedures in three patients from two centers where implantation of percutaneous valves within the scaffold of a bioprosthetic tricuspid valve was technically successful but there was early valve failure.ResultsIn all cases, immediate results suggested excellent valve function; however, valve failure occurred early in three and later in one. Two patients elected for valve explantation and surgical placement of a further tricuspid valve. The third patient underwent a second percutaneous valve-in-valve procedure though this valve is also showing early signs of failure.ConclusionsTranscatheter placement of percutaneous valves in the tricuspid position is technically a relatively straightforward procedure but early valve failure raises some concerns about the viability of the valve in this position, at least in some patients.Copyright © 2013 Wiley Periodicals, Inc.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.