    Leak Compensation During Volume Guarantee With the Dräger Babylog VN500 Neonatal Ventilator.

    • Eniko Szakmar, Colin J Morley, and Gusztav Belteki.
    • Neonatal Intensive Care Unit, The Rosie Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge, United Kingdom.
    • Pediatr Crit Care Me. 2018 Sep 1; 19 (9): 861-868.

    ObjectivesTo investigate how compensating for endotracheal tube leaks by targeting the leak-compensated tidal volume affects measured physiologic and ventilator variables during neonatal mechanical ventilation.DesignRetrospective observational study.SettingA level III Neonatal ICU.PatientsWe enrolled 30 neonates who were ventilated using synchronized intermittent positive pressure mode with volume guarantee and had at least 12 hours of continuous detailed recording of ventilation variables.InterventionsInfants were treated using the Dräger VN500 ventilator (Dräger, Lübeck, Germany), which uses a proprietary algorithm to measure and compensate for endotracheal tube leaks. Eleven were ventilated without leak compensation and 19 with leak compensation.Measurements And Main ResultsDetailed ventilation data were collected and analyzed at 1 Hz, with intermittent blood gas values. The percentage of leak was less than 20% in 73% of leak-compensated inflations, and the volume of the leak compensation was less than 1 mL/kg in 97.3% of inflations. Between the two groups, ventilation variables were comparable, except the percentage of leak that was significantly (p = 0.005) higher in the recordings with leak compensation. Without leak compensation, the mean expired tidal volume was maintained very close to the set level up to 50% leak, but with leaks greater than 50%, it declined progressively. With leak compensation, the mean leak-compensated expired tidal volume was well maintained even with leak greater than 90% although with large variability. Without leak compensation, the difference between the maximum allowed inflating pressure and the peak inflating pressure decreased progressively as the leak increased. This did not occur with leak compensation. The median PCO2 was slightly higher with leak compensation.ConclusionsDuring volume guarantee ventilation with a Dräger VN500 ventilator, without leak compensation the expired tidal volume declined after 50% leak. With leak compensation, the tidal volume was maintained even with a large leak. With leak compensation, there was a more stable peak inflating pressure, although the PCO2 was slightly higher.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…