Case ReportsValve-in-Valve Procedures in Failing Biological Xenografts Using a Novel Balloon-Expandable Device: Experience in Aortic, Mitral, and Tricuspid Positions.
- Andreas Schaefer, Lenard Conradi, Moritz Seiffert, Edith Lubos, Stefan Blankenberg, Hermann Reichenspurner, Ulrich Schaefer, and Hendrik Treede.
- Department of Cardiovascular Surgery, University Heart Center Hamburg, Hamburg, Germany.
- Thorac Cardiovasc Surg. 2016 Aug 1; 64 (5): 366-73.
AbstractBackground Valve-in-valve (ViV) procedures for degenerated bioprostheses are an alternative for the standard of care in an aging population. Several reports showed that the Edwards Sapien XT (Edwards Lifesciences Co., Irvine, California, United States) transcatheter heart valve (THV) can be used in aortic, mitral, and tricuspid position for ViV procedures. No published case series for different valve positions exist regarding suitability of the new Edwards Sapien 3 (Edwards Lifesciences Co.) THV for this purpose. Especially, the increased stent height compared with the XT and the newly added polyethylene terephthalate cuff is of potential concern in ViV interventions. Herein, we report six cases of ViV procedures with the Edwards Sapien 3 THV with a focus on technical considerations. Methods and Results Between October 2013 and November 2014, six ViV procedures with the Edwards Sapien 3 THV were performed. Four implants were done in aortic, one in mitral, and one in tricuspid position. All procedures were performed successfully without any complications. Fluoroscopy and echocardiography confirmed an adequate position and function without any paravalvular or transvalvular leakage or elevated transvalvular gradients in any case. Conclusion Preliminary experience suggests, ViV procedures with the Edwards Sapien 3 THV are safe and reliable. The outer polyethylene terephthalate cuff, for enhanced paravalvular sealing, led to a good outcome, concerning PVL in ViV procedures without resulting in elevated transvalvular gradients. This was even the case in a mildly undersized THV when compared with the internal diameter of the surgical bioprosthesis. The central radiopaque positioning marker and the fine adjustment wheel allow for accurate positioning within degenerated bioprostheses. The increased stent height, compared with the Sapien XT, led to no complications, especially in mitral position. In bioprostheses without any fluoroscopic landmarks, a balloon valvuloplasty may be necessary to identify the appropriate deployment position. Georg Thieme Verlag KG Stuttgart · New York.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.