• Br J Anaesth · Sep 2018

    Comparative Study

    Estimating the rapid haemodynamic effects of passive leg raising in critically ill patients using bioreactance.

    • L Galarza, P Mercado, J-L Teboul, V Girotto, A Beurton, C Richard, and X Monnet.
    • Service de Réanimation Médicale, Hôpital de Bicêtre, Hôpitaux Universitaires Paris-Sud, Assistance Publique - Hôpitaux de Paris, Inserm UMR S_999, Université Paris-Sud, Le Kremlin-Bicêtre, France.
    • Br J Anaesth. 2018 Sep 1; 121 (3): 567-573.

    BackgroundRapid detection of changes in cardiac index (CI) in real time using minimally invasive monitors may be of clinical benefit. We tested whether the Starling-SV bioreactance device, which averages CI over a short 8 s period, could assess the effects of passive leg raising (PLR), a clinical test that is recommended to assess fluid responsiveness during septic shock.MethodsIn 32 critically ill patients, we measured CI by transpulmonary thermodilution (PiCCO2, CItd), pulse contour analysis (PiCCO2, CIPulse), and the Starling-SV device (CIStarling) at baseline. CIPulse and CIStarling were measured again at the end of a PLR test. In the 13 patients with a positive PLR test, CItd, CIPulse, and CIStarling were measured before and after a 500 ml saline infusion. The primary outcome was relative changes from baseline measurements in CItd, CIPulse, and CIStarling. Secondary outcomes compared absolute values measured by each method.ResultsRelative changes in CIPulse and CItd were significantly correlated (r=0.82; n=45; P<0.001), with an 89% concordance rate (n=45 paired measurements). Relative changes in CIStarling and CItd were also significantly correlated (r=0.59; n=45; P<0.001) with a 78% concordance rate. For absolute measures of CI (n=77 paired measurements), the bias between CIPulse and CItd was 0.01 L min-1 m-2 (limits of agreement, -0.49 and 0.51 L min-1 m-2; 15% percentage error). Bias between CIStarling and CItd was 0.03 L min-1 m-2 (limits of agreement, -1.61 and 1.67 L min-1 m-2; 48% percentage error).ConclusionsIn critically ill patients, a non-invasive bioreactance device with a shorter averaging period assessed a passive leg raising test with reasonable accuracy.Copyright © 2018 British Journal of Anaesthesia. Published by Elsevier Ltd. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…