• Neurocritical care · Oct 2018

    Clinical Trial Observational Study

    Rapid Bedside Evaluation of Seizures in the ICU by Listening to the Sound of Brainwaves: A Prospective Observational Clinical Trial of Ceribell's Brain Stethoscope Function.

    • Kyle Hobbs, Prashanth Krishnamohan, Catherine Legault, Steve Goodman, Josef Parvizi, Kapil Gururangan, and Michael Mlynash.
    • Department of Neurology and Neurological Sciences, Stanford University School of Medicine, Stanford, CA, USA. khobbs@wakehealth.edu.
    • Neurocrit Care. 2018 Oct 1; 29 (2): 302-312.

    BackgroundPatients suffering from non-convulsive seizures experience delays in diagnosis and treatment due to limitations in acquiring and interpreting electroencephalography (EEG) data. The Ceribell EEG System offers rapid EEG acquisition and conversion of EEG signals to sound (sonification) using a proprietary algorithm. This study was designed to test the performance of this EEG system in an intensive care unit (ICU) setting and measure its impact on clinician treatment decision.MethodsEncephalopathic ICU patients at Stanford University Hospital were enrolled if clinical suspicion for seizures warranted EEG monitoring. Treating physicians rated suspicion for seizure and decided if the patient needed antiepileptic drug (AED) treatment at the time of bedside evaluation. After listening to 30 s of EEG from each hemisphere in each patient, they reevaluated their suspicion for seizure and decision for additional treatment. The EEG waveforms recorded with Ceribell EEG were subsequently analyzed by three blinded epileptologists to assess the presence or absence of seizures within and outside the sonification window. Study outcomes were EEG set up time, ease of use of the device, change in clinician seizure suspicion, and change in decision to treat with AED before and after sonification.ResultsThirty-five cases of EEG sonification were performed. Mean EEG setup time was 6 ± 3 min, and time to obtain sonified EEG was significantly faster than conventional EEG (p < 0.001). One patient had non-convulsive seizure during sonification and another had rhythmic activity that was followed by seizure shortly after sonification. Change in treatment decision after sonification occurred in approximately 40% of patients and resulted in a significant net reduction in unnecessary additional treatments (p = 0.01). Ceribell EEG System was consistently rated easy to use.ConclusionThe Ceribell EEG System enabled rapid acquisition of EEG in patients at risk for non-convulsive seizures and aided clinicians in their evaluation of encephalopathic ICU patients. The ease of use and speed of EEG acquisition and interpretation by EEG-untrained individuals has the potential to improve emergent clinical decision making by quickly detecting non-convulsive seizures in the ICU.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.