    An in vitro evaluation of the influence of neonatal endotracheal tube diameter and length on the work of breathing.

    • Fuyu Miyake, Rika Suga, Takahiro Akiyama, and Fumihiko Namba.
    • Department of Pediatrics, Saitama Medical Center, Saitama Medical University, Kawagoe, Saitama, Japan.
    • Paediatr Anaesth. 2018 May 1; 28 (5): 458-462.

    BackgroundNeonates, particularly premature babies, are often managed with endotracheal intubation and subsequent mechanical ventilation to maintain adequate pulmonary gas exchange. There is no consensus on the standard length of endotracheal tube. Although a short tube reduces resistance and respiratory dead space, it is believed to increase the risk of accidental extubation. There are not entirely coherent data regarding the effect of endotracheal tube length on work of breathing in infants.AimThe aim of this study was to evaluate the impact of neonatal endotracheal tube diameter and length on the work of breathing using an infant in vitro lung model.MethodWe assessed the work of breathing index and mechanical ventilation settings with various endotracheal tube diameters and lengths using the JTR100 in vitro infant lung model. The basic parameters of the model were breathing frequency of 20 per minutes, inspiratory-expiratory ratio of 1:3, and positive end-expiratory pressure of 5 cmH2 O. In addition, the diaphragm driving pressure to maintain the set tidal volume was measured as the work of breathing index. The JTR100 was connected to the Babylog 8000plus through the endotracheal tube. Finally, we monitored the peak inspiratory pressure generated during assist-control volume guarantee mode with a targeted tidal volume of 10-30 mL.ResultsThe diaphragm driving pressure using a 2.0-mm inner diameter tube was twice as high as that using a 4.0-mm inner diameter tube. To maintain the targeted tidal volume, a shorter tube reduced both the diaphragm driving pressure and ventilator-generated peak inspiratory pressure. The difference in the generated peak inspiratory pressure between the shortest and longest tubes was 5 cmH2 O.ConclusionIn our infant lung model, a shorter tube resulted in a lower work of breathing and lower ventilator-generated peak inspiratory pressure.© 2018 John Wiley & Sons Ltd.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.