• Paediatric anaesthesia · Sep 2018

    Randomized Controlled Trial

    Artificial ventilation during transport: A randomized crossover study of manual resuscitators with comparison to mechanical ventilators in a simulation model.

    • Malcolm J Lucy, Jonathan J Gamble, Andrew Peeling, Jimmy T H Lam, and Lloyd Balbuena.
    • Department of Anesthesiology, University of Saskatchewan, Saskatoon, SK, Canada.
    • Paediatr Anaesth. 2018 Sep 1; 28 (9): 788-794.

    BackgroundPositive-pressure ventilation in critically ill patients is commonly administered via a manual resuscitation device or a mechanical ventilator during transport. Our group previously compared delivered ventilation parameters between a self-inflating resuscitator and a flow-inflating resuscitator during simulated in-hospital pediatric transport. However, unequal group access to inline pressure manometry may have biased our results. In this study, we examined the performance of the self-inflating resuscitator and the flow-inflating resuscitator, both equipped with inline manometry, and several mechanical ventilators to deliver prescribed ventilation parameters during simulated pediatric transport.MethodsThirty anesthesia providers were randomized to initial resuscitator device used to hand ventilate a test lung. The resuscitators studied were a Jackson-Rees circuit (flow-inflating resuscitator) or a Laerdal pediatric silicone resuscitator (self-inflating resuscitator), both employing manometers. The scenario was repeated using several mechanical transport ventilators (Hamilton-T1, LTV® 1000, and LTV® 1200). The primary outcome was the proportion of total breaths delivered within the predefined target PIP/PEEP range (30 ± 3, 10 ± 3 cm H2 O).ResultsThe Hamilton-T1 outperformed the other ventilators for breaths in the recommended range (χ2  = 2284, df = 2, P < .001) and with no breaths in the unacceptable range (χ2  = 2333, df = 2, P < .001). Hamilton-T1 also outperformed all human providers in proportion of delivered acceptable and unacceptable breaths (χ2  = 4540, df = 3, P < .001 and χ2  = 639, df = 3, P < .001, respectively). Compared with the flow-inflating resuscitator, the self-inflating resuscitator was associated with greater odds of breaths falling outside the recommended range (Odds ratio (95% CI): 1.81 (1.51-2.17)) or unacceptable (Odds ratio (95% CI): 1.63 (1.48-1.81)).ConclusionThis study demonstrates that a majority of breaths delivered by manual resuscitation device fall outside of target range regardless of provider experience or device type. The mechanical ventilator (Hamilton-T1) outperforms the other positive-pressure ventilation methods with respect to delivery of important ventilation parameters. In contrast, 100% of breaths delivered by the LTV 1200 were deemed unacceptable.© 2018 John Wiley & Sons Ltd.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.