    Evaluation of the Fourth-Generation FloTrac/Vigileo System in Comparison With the Intermittent Bolus Thermodilution Method in Patients Undergoing Cardiac Surgery.

    • Yusuke Kusaka, Fumihiro Ohchi, and Toshiaki Minami.
    • Department of Anesthesiology, Osaka Medical College, Takatsuki, Osaka, Japan. Electronic address: ane063@poh.osaka-med.ac.jp.
    • J. Cardiothorac. Vasc. Anesth. 2019 Apr 1; 33 (4): 953-960.

    ObjectivesThe aim of this study was to evaluate the accuracy, precision, and trending ability of the fourth-generation FloTrac/Vigileo system (version 4.00; Edwards Lifesciences, Irvine, CA) by comparing cardiac output derived from FloTrac/Vigileo system (COAP) with that measured by a pulmonary artery catheter (COTD), and to determine the effects of hemodynamic variables on the bias between COTD and COAP.DesignA prospective study.SettingUniversity hospital.ParticipantsThirty patients undergoing elective cardiac surgery using cardiopulmonary bypass.InterventionsIncluding hemodynamic variables, COTD and COAP were measured simultaneously at the following 10 time points: after the induction of anesthesia, at the start of operation, after sternotomy, before and after the administration of heparin, before and after the administration of protamine, at the start of sternal closure, at the end of operation, and on arrival to intensive care unit.Measurements And Main ResultsIn total, 280 pairs of datasets were obtained. Bland-Altman analysis showed a bias of -0.41 L/min, a precision of 0.72 L/min, and limits of agreement of -1.85 and 1.03 L/min, with a percentage error of 37.1%. The concordance rate determined by 4-quadrant plot analysis and the polar concordance rate were 76% and 79%, respectively. The linear mixed-effect model revealed that the bias was influenced strongly by the difference in pulse pressure between the radial and femoral artery (p < 0.001), and the systemic vascular resistance index (p < 0.001).ConclusionThe fourth-generation FloTrac/Vigileo system still lacks accuracy and trending ability in cardiac surgery, and the discrepancy in cardiac output measurement depends on the peripheral vascular tone. Further improvement of this system is needed.Copyright © 2018 Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…