Validation of inflationary noninvasive blood pressure monitoring in the emergency room.
- Junichi Sasaki, Yoshiharu Kikuchi, Takashi Usuda, and Shingo Hori.
- aDepartment of Emergency and Critical Care Medicine, Keio University School of Medicine bNihon Kohden Corporation, Tokyo, Japan.
- Blood Press Monit. 2015 Dec 1; 20 (6): 325-9.
ObjectiveThe aim of this study was to compare the values of a noninvasive blood pressure (NIBP) measurement during cuff inflation (inflationary NIBP) and deflationary NIBP measurements and to verify whether inflationary NIBP is equivalent to conventional deflationary NIBP and is acceptable for clinical use in the emergency room (ER).Materials And MethodsA total of 2981 NIBP data points were collected from 175 patients (age, 56.5±22.2 years; range, 7-92 years) who had been treated in the resuscitation area of the ER at Keio University Hospital. The data points were obtained using two alternate algorithms with a standard monitor (BSM-6000). One algorithm consisted of continuous inflationary and deflationary measurements in a single cycle (dual algorithm, 1502 data points); this algorithm was used to verify the success rate and the precision of the data. The second algorithm (1479 data points) consisted of only conventional deflationary measurements and was used to verify the duration of the measurement cycle.ResultsThe success rate of the inflationary NIBP (completed using only the inflationary method) was 69.0%. Failures in the inflationary measurements were caused by arrhythmia and/or body motions. The mean difference and SD of the systolic pressure and the diastolic pressure between inflationary NIBP and deflationary NIBP were -0.6±8.8 and 3.5±7.5 mmHg, respectively. The confidence intervals were -0.6 (95% confidence interval=-1.1 to -0.1) and 3.5 (95% confidence interval=3.0 to 4.0) mmHg. The coefficients of correlation were 0.96 and 0.93. Inflationary NIBP was capable of determining the NIBP more quickly compared with deflationary NIBP (average of 15.9 vs. 34.2 s; P<0.05).ConclusionInflationary NIBP measurements have a reasonable accuracy and a sufficient rapidity, compared with deflationary NIBP measurements, in ER patients.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.