- Christanne Coffey, John Serra, Mat Goebel, Sarah Espinoza, Edward Castillo, and James Dunford.
- University of California San Diego Health System, San Diego, California.
- J Emerg Med. 2018 Jul 1; 55 (1): 71-77.
BackgroundA significant increase in false positive ST-elevation myocardial infarction (STEMI) electrocardiogram interpretations was noted after replacement of all of the City of San Diego's 110 monitor-defibrillator units with a new brand. These concerns were brought to the manufacturer and a revised interpretive algorithm was implemented.ObjectivesThis study evaluated the effects of a revised interpretation algorithm to identify STEMI when used by San Diego paramedics.MethodsData were reviewed 6 months before and 6 months after the introduction of a revised interpretation algorithm. True-positive and false-positive interpretations were identified. Factors contributing to an incorrect interpretation were assessed and patient demographics were collected.ResultsA total of 372 (234 preimplementation, 138 postimplementation) cases met inclusion criteria. There was a significant reduction in false positive STEMI (150 preimplementation, 40 postimplementation; p < 0.001) after implementation. The most common factors resulting in false positive before implementation were right bundle branch block, left bundle branch block, and atrial fibrillation. The new algorithm corrected for these misinterpretations with most postimplementation false positives attributed to benign early repolarization and poor data quality. Subsequent follow-up at 10 months showed maintenance of the observed reduction in false positives.ConclusionsThis study shows that introducing a revised 12-lead interpretive algorithm resulted in a significant reduction in the number of false positive STEMI electrocardiogram interpretations in a large urban emergency medical services system. Rigorous testing and standardization of new interpretative software is recommended before introduction into a clinical setting to prevent issues resulting from inappropriate cardiac catheterization laboratory activations.Copyright © 2018 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.