Reg Anesth Pain Med · Jan 2019
Evaluation of the B-Smart manometer and the CompuFlo computerized injection pump technology for accurate needle-tip injection pressure measurement during peripheral nerve blockade.
- Robyn S Weisman, Nirav P Bhavsar, Kathleen A Schuster, and Ralf E Gebhard.
- Department of Anesthesiology, University of Miami - Miller School of Medicine, Miami, Florida, USA rweisman1@med.miami.edu.
- Reg Anesth Pain Med. 2019 Jan 1; 44 (1): 86-90.
Background And ObjectivesThe exact mechanism of peripheral nerve blocks causing/leading to nerve injury remains controversial. Evidence from animal experiments suggests that intrafascicular injection resulting in high injection pressure has the potential to rupture nerve fascicles and may consequently cause permanent nerve injury and neurological deficits. The B-Smart (BS) in-line manometer and the CompuFlo (CF) computerized injection pump technology are two modalities used for monitoring pressure during regional anesthesia. This study sought to explore the accuracy of these two technologies in measuring needle-tip pressures in a simulated environment.MethodsIn seven simulated needle-syringe combinations, the BS and the CF devices were connected in series through a closed system and attached to a digital manometer at the tip of various needles. The pressures were evaluated in three trials per needle-syringe combination. Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy (F1 Score) were determined for each needle type and overall.ResultsFor pressures ≥15 psi and ≥20 psi, respectively, the CF device demonstrated a sensitivity of 100%, 100%; specificity of 96%, 98%; positive predictive value 93%, 93%; and negative predictive value of 100%, 100%. The BS device demonstrated a sensitivity of 60%, 100%; specificity of 99%, 95%; positive predictive value of 96%, 85%; and negative predictive value of 85%, 100%. Accuracy, as measured by the F1 Score, for detecting a pressure of ≥15 psi was 0.96 for the CF and 0.74 for the BS.ConclusionsFuture research is needed to explore in-vivo performance and evaluate whether either of these devices can impact on clinical outcomes.© American Society of Regional Anesthesia & Pain Medicine 2019. No commercial re-use. See rights and permissions. Published by BMJ.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.