• Respiratory care · Sep 2018

    Comparative Study

    The Effect of Different Closed Suction Catheter Designs and pMDI Adapters on Aerosol Delivery in Simulated Adult Mechanical Ventilation With and Without Exhaled Humidity.

    • Jacqueline P Williams, Arzu Ari, Ramalingam Shanmugam, and James B Fink.
    • Respiratory Therapy Department, Emory University Midtown Hospital, Emory Healthcare, Atlanta, Georgia.
    • Respir Care. 2018 Sep 1; 63 (9): 1154-1161.

    BackgroundClosed suction catheters are commonly used with pressurized metered-dose inhalers (pMDIs) in ventilator-dependent patients receiving bronchodilator therapy. Although heated/humidified circuits result in a decrease in drug delivery, no studies have investigated the impact of a closed suction catheter with a pMDI on aerosol delivery during mechanical ventilation. We sought to quantify aerosol delivery with a variety of closed suction catheters and pMDI adapters in a simulated ventilator-dependent adult lung model with and without exhaled humidity.MethodsA ventilator (tidal volume 450 mL, PEEP 5 cm H2O, breathing frequency 15 breaths/min, peak flow 60 L/min, and bias flow 2 L/min) was attached to 3 closed suction catheter designs (T-piece, double swivel elbow, and a multi-access port) with pMDI adapters (small- and large-unidirectional adapters, a bidirectional adapter, and a built-in port) connected to an endotracheal tube (ETT) with a collecting filter and a passive test lung. To simulate exhaled humidity, a heated humidifier was placed between the collection filter and the distal tip of the ETT, set at 36°C. Experiments without exhaled humidity were performed with no heat and humidification. For each test, 4 puffs of albuterol sulfate (432 μg) were administered. The drug was eluted from the collecting filter and analyzed via spectrophotometer (276 nm). Each experiment was run 3 times.ResultsDelivery efficiency of the bidirectional adapter was greatest compared to other adapters tested in this study (P = .01). Aerosol deposition obtained from the combination of the T-piece closed suction catheter and the bidirectional adapter was more than that from the multi-access port (P = .42) and double swivel closed suction catheter (P = .27). Drug delivery without exhaled humidity exceeded values with exhaled humidity by 20-90% using the small unidirectional, large unidirectional, and bidirectional adapters (P = .01, P = .02 and P = .02, respectively).ConclusionsClosed suction catheter, pMDI spacer/adapters, and exhalation conditions have an impact on drug delivery during simulated adult mechanical ventilation. Aerosol delivery without exhaled humidity appears to report unrealistically high values in comparison to the more accurate values with simulated exhalation humidity.Copyright © 2018 by Daedalus Enterprises.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…