    Size matters: An observational study investigating estimated height as a reference size for calculating tidal volumes if low tidal volume ventilation is required.

    • Benjamin Sasko, Ulrich Thiem, Martin Christ, Hans-Joachim Trappe, Oliver Ritter, and Nikolaos Pagonas.
    • Department of Cardiology, Brandenburg Medical School, University Hospital Brandenburg, Germany.
    • Plos One. 2018 Jan 1; 13 (6): e0199917.

    PurposeAcute lung injury is a life threatening condition often requiring mechanical ventilation. Lung-protective ventilation with tidal volumes of 6 mL/kg predicted body weight (PBW, calculated on the basis of a patient's sex and height), is part of current recommended ventilation strategy. Hence, an exact height is necessary to provide optimal mechanical ventilation. However, it is a common practice to visually estimate the body height of mechanically ventilated patients and use these estimates as a reference size for ventilator settings. We aimed to determine if the common practice of estimating visual height to define tidal volume reduces the possibility of receiving lung-protective ventilation.MethodsIn this prospective observational study, 28 mechanically ventilated patients had their heights visually estimated by 20 nurses and 20 physicians. All medical professionals calculated the PBW and a corresponding tidal volume with 6 ml/kg/PBW on the basis of their visual estimation. The patients' true heights were measured and the true PBW with a corresponding tidal volume was calculated. Finally, estimates and measurements were compared.Results1033 estimations were undertaken by 153 medical professionals. The majority of the estimates were imprecise and resulting data comprised taller body heights, higher PBW and higher tidal volumes (all p≤0.01). When estimates of patients´ heights are used as a reference for tidal-volume definition, patients are exposed to mean tidal volumes of 6.5 ± 0.4 ml/kg/PBW. 526 estimation-based tidal volumes (51.1%) did not provide lung-protective ventilation. Shorter subjects (<175cm) were a specific risk group with an increased risk of not receiving lung protective ventilation (OR 6.6; 95%CI 1.2-35.4; p = 0.02), while taller subjects had a smaller risk of being exposed to inadequately high tidal volumes (OR 0.15; 95%CI 0.02-0.8; p = 0.02). Furthermore, we found an increased risk of overestimating if the assessor was a female (OR 1.74; 95%CI 1.14-2.65; p = 0.01).ConclusionThe common practice of visually estimating body height and using these estimates for ventilator settings is imprecise and potentially harmful because it reduces the chance of receiving lung-protective ventilation. Avoiding this practice increases the patient safety. Instead, height should be measured as a standard procedure.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…