• J Orthop Sci · Jan 1999

    Case Reports

    Accurate pedicle screw insertion under the control of a computer-assisted image guiding system: laboratory test and clinical study.

    • M Kamimura, S Ebara, H Itoh, Y Tateiwa, T Kinoshita, and K Takaoka.
    • Shinshu University School of Medicine, Department of Orthopaedic Surgery, 3-1-1 Asahi Matsumoto, Nagano 390-0802, Japan.
    • J Orthop Sci. 1999 Jan 1; 4 (3): 197-206.

    AbstractWe used a commercially available computer-assisted navigation system (StealthStation; Sofamor Danek, Memphis, TN, USA) in both an in-vitro and a clinical study performed in 1996-1998. The basic data used for navigation were preoperative computed tomography (CT) scan imaging data. The position of the probe or drill guide was superimposed in real-time on a monitor. For the in-vitro study, ten plastic lumbar spine models (50 vertebrae) were used. The entrance hole for the screw was made by drilling, following navigation. Using the navigation system, we drilled 88 holes through the pedicles into the vertebral bodies of 44 vertebral models. All 88 pedicle holes were contained within the pedicle without perforation. The mean deviation of the hole positions from the surgical plan was 1.78 +/- 0.81 mm, and the mean angular deviation was 2.28 degrees +/- 1.92 degrees. In 29 patients, using the navigation system, we introduced 169 pedicle screws at the planned position. Fifty-one screws were used for thoracic and 118 screws for lumbar spinal fixation. All screws correctly passed through the pedicles. There were no neurological complications after surgery. Using this guided surgery system, we achieved satisfactory results both in the laboratory and in a clinical setting.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.