- A van Drumpt, J van Bommel, S Hoeks, F Grüne, T Wolvetang, J Bekkers, and M Ter Horst.
- Department of Anesthesiology, Erasmus Medical Center, PO Box 2040, 3000 CA, Rotterdam, The Netherlands. annevandrumpt@hotmail.com.
- BMC Anesthesiol. 2017 Mar 14; 17 (1): 42.
BackgroundA relatively new uncalibrated arterial pressure waveform cardiac output (CO) measurement technique is the Pulsioflex-ProAQT® system. Aim of this study was to validate this system in cardiac surgery patients with a specific focus on the evaluation of a difference in the radial versus the femoral arterial access, the value of the auto-calibration modus and the ability to show fluid-induced changes.MethodsIn twenty-five patients scheduled for ascending aorta, aortic arch replacement, or both we measured CO simultaneously by transpulmonary thermodilution (COtd) and by using the ProAQT® system connected to the radial (COpR), as well as the femoral artery catheter (COpF). Hemodynamic data were assessed at predefined time points; from incision until 16 h after ICU admission.ResultsIn total 175 (radial) and 179 (femoral) pairs of CO measurement were collected. The accuracy of COpR/COpF was evaluated showing a mean bias of -0.31 L/min (±2.9 L/min) and -0.57 L/min (± 2.8 L/min) with percentage errors of 49 and 46% respectively. Trending ability of the ProAQT® device was evaluated; the four quadrant concordance rates in the radial and femoral artery were 74 and 75% and improved to 77 and 85% after auto-calibration. The mean angular biases in the radial and femoral artery were 6.4° and 6.0° and improved to 5° and 3.3° after auto-calibration. The polar concordance rates in the radial and femoral artery were 65 and 70% and improved to 76 and 84% after auto-calibration. Considering the fluid-induced changes in stroke volume(SV), the coefficient of correlation between the changes in SVtd and SVp was 0.57 (p < 0.01) in the radial artery and 0.60 (p < 0.01) in the femoral artery.ConclusionsThe ProAQT® system can be of additional value if the clinician wants to determine fluid responsiveness in cardiac surgery patients. However, the ProAQT® system provided inaccurate CO measurements compared to transpulmonary thermodilution. The trending ability was poor for COpR but moderate for COpF. Auto-calibration of the system did not improve accuracy of CO measurements nor did it improve the prediction of fluid responsiveness. However, the trending ability was improved by auto-calibration, possibly by correcting a drift over a longer time period.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.