• Stroke · Mar 2016

    Flow Diverter Therapy With the Pipeline Embolization Device Is Associated With an Elevated Rate of Delayed Fluid-Attenuated Inversion Recovery Lesions.

    • Mina G Safain, Marie Roguski, Robert S Heller, and Adel M Malek.
    • From the Cerebrovascular and Endovascular Division, Department of Neurosurgery, Tufts Medical Center and Tufts University School of Medicine, Boston, MA.
    • Stroke. 2016 Mar 1; 47 (3): 789-97.

    Background And PurposeFlow diversion using the Pipeline Embolization Device is reported as a safe treatment of aneurysms. Complete aneurysm occlusion, however, occurs in a delayed fashion with initial persistent filling of the aneurysm dome. We hypothesized that this transflow across metallic struts may be associated with thromboembolic events.MethodsForty-one consecutive patients undergoing aneurysm treatment with the Pipeline Embolization Device and a comparison group of 78 Neuroform stent-mediated embolizations were studied. Patients' charts, procedure notes, platelet function, and anticoagulation state were analyzed. Serial magnetic resonance images were assessed for the presence of newly occurring diffusion-weighted imaging and fluid-attenuated inversion recovery (FLAIR) lesions at multiple postprocedure time ranges (average days post procedure [Pipeline Embolization Device/Neuroform]: T1=1, T2=73/107, T3=174, T4=277/335, and T5=409). In addition, diffusion-weighted imaging or FLAIR burden was estimated by lesional diameter summation.ResultsPipeline patients were more likely to have new ipsilateral FLAIR lesions at all time points studied (30.6% versus 7.2% of patients at T=2 and 34.5% versus 6.2% at T=4). The mean FLAIR burden was significantly increased for Pipeline patients (10.1 versus 0.7 mm at T=2 and 8.8 versus 1.9 mm at T=4). Overall 34% (14/41) of Pipeline patients experienced a new FLAIR lesion at anytime when compared with 10% (8/78) of Neuroform stent-coil patients. Postprocedural diffusion-weighted imaging did not predict future FLAIR lesions suggesting a nonprocedural cause.ConclusionsThe Pipeline Embolization Device is associated with increased rate of de novo FLAIR lesions occurring in a delayed fashion and distinct from perioperative diffusion-weighted imaging lesions. The cause and clinical effect of these lesions are unknown and suggest the need for prudent follow-up and evaluation.© 2016 American Heart Association, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…