    Optimization of the circuit configuration of a pulsatile ECLs: an in vivo experimental study.

    • Choon Hak Lim, Ho Sung Son, Jung Joo Lee, Yong Hu Fang, Ki Chul Moon, Chi Bum Ahn, Kyung Hyun Kim, Hye Won Lee, and Kyung Sun.
    • Department of Anesthesiology and Pain Medicine, Korea University, Seoul, Korea
    • ASAIO J. 2005 Sep 1; 51 (5): 609-13.

    AbstractAn extracorporeal life support system (ECLS) with a conventional membrane oxygenator requires a driving force for the blood to pass through hollow fiber membranes. We hypothesized that if a gravity-flow hollow fiber membrane oxygenator is installed in the circuit, the twin blood sacs of a pulsatile ECLS (the Twin-Pulse Life Support, T-PLS) can be placed downstream of the membrane oxygenator. This would increase pump output by doubling pulse rate at a given pumpsetting rate while maintaining effective pulsatility. The purpose of this study was to determine the optimal circuit configuration for T-PLS with respect to energy and pump output. Animals were randomly assigned to 2 groups in a total cardiopulmonary bypass model. In the serial group, a conventional membrane oxygenator was located between the twin blood sacs of the T-PLS. In the parallel group, the twin blood sacs were placed downstream of the gravity-flow membrane oxygenator. Energy equivalent pressure (EEP), surplus hemodynamic energy (SHE) and pump output were collected at the different pump-setting rates of 30, 40, and 50 beats per minute (BPM). At a given pump-setting rate the pulse rate doubled in the parallel group. Percent changes of mean arterial pressure to EEP were 13.0 +/- 1.7, 12.0 +/- 1.9, and 7.6 +/- 0.9% in the parallel group, while 22.5 +/- 2.4, 23.2 +/- 1.9, and 21.8 +/- 1.4 in the serial group at 30, 40, and 50 BPM of pump-setting rates. SHE at each pump setting rate was 20,131 +/- 1408, 21,739 +/- 2470, and 15,048 +/- 2108 erg/ cm3 in the parallel group, while 33,968 +/- 3001, 38,232 +/- 3281, 37,964 +/- 2693 erg/cm3 in the serial group. Pump output was higher in the parallel circuit at 40, and 50 BPM pump-setting rates (3.1 +/- 0.2, 3.7 +/- 0.2 L/min vs. 2.2 +/- 0.1 and 2.5 +/- 0.1 L/min, respectively, p =0.01). Either parallel or serial circuit configuration of T-PLS generates effective pulsatility. As for the pump out, the parallel circuit configuration provides higher flow than the serial circuit configuration by doubling the pulse rate at a given pump-setting rate.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.