• Neurosurgery · Jan 2005

    Comparative Study

    Biomechanical comparison of two stabilization techniques of the atlantoaxial joints: transarticular screw fixation versus screw and rod fixation.

    • Hiroshi Kuroki, Setti S Rengachary, Vijay K Goel, Scott A Holekamp, Ville Pitkänen, and Nabil A Ebraheim.
    • Department of Orthopedic Surgery, University of Miyazaki, Faculty of Medicine, Miyazaki, Japan. hiroshik@med.miyazaki-u.ac.jp
    • Neurosurgery. 2005 Jan 1; 56 (1 Suppl): 151-9; discussion 151-9.

    ObjectiveTo compare the biomechanical stability imparted to the C1 and C2 vertebrae by either transarticular screw fixation (TSF) or screw and rod fixation (SRF) techniques in a cadaver model.MethodsTen fresh ligamentous human cervical spine specimens were harvested from cadavers. The specimens were tested sequentially in the intact state, after injury and stabilization (unilateral left side and bilateral), and after fatiguing to 5000 cycles (0.5 Hz) at +/-1.0 N.m of flexion and extension. The specimens were stabilized by use of TSF in 5 spines or SRF in the other 5 spines. The data were converted to angular displacements, and the stabilized cases were compared with intact states for evaluating the efficacies of the two techniques in stabilizing the C1-C2 segments.ResultsIn the TSF group, the unilateral fixation using one screw imparted a significant stability in only the axial rotation mode. The unilateral procedure in the SRF group was effective in stabilization in all modes except in extension. The bilateral procedure in both of the groups was effective across the C1-C2 segment. However, the SRF group afforded higher stability than the corresponding TSF group in the flexion and extension modes. The degree of stability did not change after fatigue compared with the prefatigue data.ConclusionIn general, a surgeon should undertake a bilateral fixation to achieve sufficient stability across the atlantoaxial complex, and either technique will provide satisfactory results, although the SRF technique may be better in the flexion and extension modes. One should use the SRF procedure while trying to achieve stability with a unilateral system.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.