• Neurosurgery · Oct 2005

    Comparative Study

    Comparison of magnetic tracking and optical tracking by simultaneous use of two independent frameless stereotactic systems.

    • Christopher R Mascott.
    • Department of Neurosurgery, Royal College of Surgeons of Ireland, Dublin, Ireland. MascottDonnelly@netscape.net
    • Neurosurgery. 2005 Oct 1; 57 (4 Suppl): 295-301; discussion 295-301.

    ObjectiveThe goal of this study was to compare the calculated accuracy and the true surgical accuracy of a magnetic and an optical tracking system at surgical registration and during surgery.MethodsTwo Food and Drug Administration-approved, commercially available frameless stereotactic systems were used simultaneously in 70 surgical cases. The Compass Cygnus-PFS system (Compass International, Rochester MN) uses magnetic field referencing and tracking. The StealthStation (Medtronic SNT, Louisville, CO) uses optical referencing and tracking. Registration was performed for each of the systems using adhesive fiducial markers, cranial-implanted markers, anatomic landmarks, or a combination thereof. Preoperative imaging consisted of volumetric computed tomography, magnetic resonance imaging, or both. Calculated accuracy was given by each of the systems as the root mean square after registration. Surgical accuracy was assessed by comparing the anatomic accuracy of each system with a number of recognizable intraoperative anatomic landmarks.ResultsCalculated accuracy (root mean square) was 1.4 +/- 0.6 mm using the magnetic system and 1.4 +/- 0.8 mm using the optical tracking system. In the 42 patients with implanted cranial fiducials, the calculated accuracies were 1.0 +/- 0.5 mm (magnetic) and 0.9 +/- 0.4 mm (optical). True surgical accuracy was considered good (3 mm or less) in both systems in 60 of 70 patients. In two patients, neither system was accurate. In eight patients, one of the two systems was considered inaccurate. Of these, the magnetic system was considered inaccurate three times and the optical system five times.ConclusionMagnetic referencing and tracking was found to be comparable with optical tracking both with regard to calculated and true surgical accuracy. Interference from metal objects in the magnetic field was seen rarely.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.