• Neurosurgery · Oct 2005

    Comparative Study

    Reduction of magnetic resonance imaging-related heating in deep brain stimulation leads using a lead management device.

    • Kenneth B Baker, Jean Tkach, John D Hall, John A Nyenhuis, Frank G Shellock, and Ali R Rezai.
    • Center for Neurological Restoration, The Cleveland Clinic Foundation, Cleveland, Ohio, USA. bakerk@ccf.org
    • Neurosurgery. 2005 Oct 1; 57 (4 Suppl): 392-7; discussion 392-7.

    ObjectiveTo evaluate the ability of a lead management device to reduce magnetic resonance imaging (MRI)-related heating of deep brain stimulation (DBS) leads and thereby to decrease the risks of exposing patients with these implants to MRI procedures.MethodsExperiments were performed using the Activa series (Medtronic, Inc., Minneapolis, MN) DBS systems in an in vitro, gelled-saline head and torso phantom. Temperature change was recorded using fluoroptic thermometry during MRI performed using a transmit-and-receive radiofrequency body coil at 1.5 T and a transmit-and-receive radiofrequency head coil at 3 T. A cranial model placed in the phantom was used to test a custom-designed burr hole device that permitted the placement of small-diameter, concentric loops around the burr hole at the DBS lead as it exited the cranium.ResultsA total of 41 scans were performed, with absolute temperature changes ranging from 0.8 to 10.3 degrees C. Depending on the MRI system tested and the side of the phantom on which the hardware was placed, loop placement resulted in reductions in temperature rise of 41 to 74%. The effect was linearly related to the number of loops formed (P < 0.01) over the range tested (0-2.75 loops).ConclusionSmall, concentric loops placed around the burr hole seem to reduce MRI-related heating for these implants. Although the mechanism is still not fully understood, a device such as that used in the present study could permit a wider range of clinical scanning sequences to be used at 1.5 and 3 T in patients with DBS implants, in addition to increasing the margin of safety for the patient.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…