J Clin Monit Comput · Dec 2018
Comparative StudyA novel device for air removal from vascular access line: a bench study.
- Zeping Xu, Lingyan Jin, Byron Smith, Yiping Bai, Haoxiang Luo, Lars Axel Strombergsson, Min Fei, and Yandong Jiang.
- Department of Anesthesiology, Jiangsu Cancer Hospital, Nanjing, China.
- J Clin Monit Comput. 2018 Dec 1; 32 (6): 1041-1047.
AbstractEfficient air removal from a vascular access line is a key step to prevent air embolism. Existing devices, especially for rapid infusers, are far from optimum. In this study, we developed a novel device, vascular access line air removal device (VALARD), and compared its efficiency of air removal and pause time of forward bulk flow with a commonly used device, the Belmont pump. Part I experiment, saline was infused at a forward bulk flow rate of 250, 500, and 750 mL/min. Meanwhile, air was introduced into the infusion line at a rate of 5, 10, and 15 mL/min for each bulk flow rate. Air bubbles > 10 µL downstream from either the VALARD or the Belmont pump and the fraction of pause time of the forward bulk flow were determined. Part II experiment, 120 mL of air was rapidly introduced into the VALARD at a bulk flow rate of about 500 mL/min. Air bubbles > 10 µL downstream from the VALARD, fraction of pause time of the forward bulk flow, and the transit time of the 120 mL of air at the working chamber were recorded. The VALARD: no air bubbles > 10 µL were detected during any tested combination of air injection and bulk flow rates without pause of forward flow. The Belmont pump: air bubbles > 10 µL were detected in 60% of the tests with pause of the forward flow. The VALARD eliminates air efficiently without pause of the forward bulk flow. Further clinical trials are needed to compare the VALARD with other devices and to assess its efficiency, safety, and user friendliness.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.