-
J. Am. Coll. Cardiol. · Mar 2014
Multicenter StudyOff-pump transapical implantation of artificial neo-chordae to correct mitral regurgitation: the TACT Trial (Transapical Artificial Chordae Tendinae) proof of concept.
- Joerg Seeburger, Mauro Rinaldi, Sten Lyager Nielsen, Stefano Salizzoni, Ruediger Lange, Markus Schoenburg, Ottavio Alfieri, Michael Andrew Borger, Friedrich Wilhelm Mohr, and Audrius Aidietis.
- Department of Cardiac Surgery, Heartcentre Leipzig, Leipzig, Germany. Electronic address: joerg.seeburger@medizin.uni-leipzig.de.
- J. Am. Coll. Cardiol. 2014 Mar 11; 63 (9): 914-9.
ObjectivesThe goal of this study was to evaluate the safety and performance of the NeoChord DS1000 system (NeoChord, Inc., Minneapolis, Minnesota).BackgroundThere is an increasing interest in transcatheter mitral valve (MV) treatment. The NeoChord DS 1000 system enables off-pump beating heart transapical MV repair with implantation of artificial neo-chordae.MethodsPatients with severe mitral regurgitation (MR) due to isolated posterior prolapse were included in this TACT (Transapical Artificial Chordae Tendinae) trial. All patients were scheduled for off-pump transapical implantation of neo-chordae.ResultsThirty patients at 7 centers were enrolled. Major adverse events included 1 death due to post-cardiotomy syndrome and concomitant sepsis and 1 minor stroke with the patient fully recovered at the 30-day follow-up visit. Additional patients experienced procedural major adverse events related to a reoperation or conversion to standard of care. Acute procedural success (placement of at least 1 neo-chord and reduction of MR from 3+ or 4+ to ≤2+) was achieved in 26 patients (86.7%). In 4 patients neo-chordae were not placed for technical and/or patient-specific reasons. These patients underwent intraoperative (3 patients) or post-operative (1 patient) standard MV repair. At 30 days, 17 patients maintained an MR grade ≤2+. Four patients who developed recurrent MR were successfully treated with open MV repair during 30-day follow-up. Results improved with experience: durable reduction in MR to ≤2+ at 30 days was achieved in 5 (33.3%) of the first 15 patients and 12 (85.7%) of the last 14 patients.ConclusionsOff-pump transapical implantation of artificial chordae to correct MR is technically safe and feasible; however, it yields further potential for improvement of efficacy and durability. (Safety and Performance Study of the NeoChord Device [TACT]; NCT01777815).Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.