    Accuracy of blood glucose measurements using the NOVA StatStrip® glucometer during cardiac surgery: a prospective observational study.

    • Yosuke Nakadate, Hiroaki Sato, Patricia Roque, Tamaki Sato, Takashi Matsukawa, Linda Wykes, Akiko Kawakami, and Thomas Schricker.
    • Department of Anesthesiology, University of Yamanashi, 1110 Shimokato, Chuo-city, Yamanashi, 409-3898, Japan. ynakadate@yamanashi.ac.jp.
    • Can J Anaesth. 2019 Aug 1; 66 (8): 943-952.

    PurposeThe Nova StatStrip® Glucose Hospital Meter System (Nova Biomedical, Waltham, MA, USA) is United States Food and Drug Administration approved for point-of-care use in critically ill patients, but its use during cardiac surgery has not been evaluated. In this study, we compare glucose values obtained during cardiac surgery by StatStrip® with values obtained by a blood gas analyzer.MethodsBlood glucose concentrations were analyzed in 121 patients by the StatStrip point- of-care test (POCT) glucose monitor and the GEM® Premier™ 3000 blood gas analyzer (Instrumentation Laboratory Company, Bedford MA, USA). Arterial blood samples were taken at baseline (before surgery), before cardiopulmonary bypass (CPB), during early and late CPB, and 30 min after CPB. Accuracy of the StatStrip glucometer was analyzed using the Clinical and Laboratory Standards Institute (CLSI) POCT12-A3 criteria (criterion 1; 95% of samples should be ± 0.66 mMol·L-1 of reference glucose values < 5.5 mMol·L-1 and ± 12.5% for reference glucose values > 5.5 mMol·L-1, criterion 2; 98% of samples should be ± 0.83 mMol·L-1 of reference glucose values < 4.1 mMol·L-1 or 20% of the reference glucose for values > 4.1 mMol·L-1).ResultsThe accuracy of StatStrip glucose measurements at baseline (99%, 100%) and before CPB (95%, 98%), but not during (early: 84%, 97%; late: 83%, 96%) and after (92%, 100%) CPB, satisfied the CLSI POCT12-A3 criteria.ConclusionArterial blood glucose measurement by StatStrip was accurate before CPB, but lacked accuracy during and after CPB. Glucose values should be interpreted with caution when intensive glucose control protocols are being used during cardiac surgery.Trial RegistrationClinicalTrials.gov (NCT02729064); registered 5 April, 2016.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…