Observational Study
Augmented Reality Surgical Navigation in Spine Surgery to Minimize Staff Radiation Exposure.
- Erik Edström, Gustav Burström, Artur Omar, Rami Nachabe, Michael Söderman, Oscar Persson, Paul Gerdhem, and Adrian Elmi-Terander.
- Department of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.
- Spine. 2020 Jan 1; 45 (1): E45-E53.
Study DesignProspective observational study.ObjectiveTo assess staff and patient radiation exposure during augmented reality surgical navigation in spine surgery.Summary Of Background DataSurgical navigation in combination with intraoperative three-dimensional imaging has been shown to significantly increase the clinical accuracy of pedicle screw placement. Although this technique may increase the total radiation exposure compared with fluoroscopy, the occupational exposure can be minimized, as navigation is radiation free and staff can be positioned behind protective shielding during three-dimensional imaging. The patient radiation exposure during treatment and verification of pedicle screw positions can also be reduced.MethodsTwenty patients undergoing spine surgery with pedicle screw placement were included in the study. The staff radiation exposure was measured using real-time active personnel dosimeters and was further compared with measurements using a reference dosimeter attached to the C-arm (i.e., a worst-case staff exposure situation). The patient radiation exposures were recorded, and effective doses (ED) were determined.ResultsThe average staff exposure per procedure was 0.21 ± 0.06 μSv. The average staff-to-reference dose ratio per procedure was 0.05% and decreased to less than 0.01% after a few procedures had been performed. The average patient ED was 15.8 ± 1.8 mSv which mainly correlated with the number of vertebrae treated and the number of cone-beam computed tomography acquisitions performed. A low-dose protocol used for the final 10 procedures yielded a 32% ED reduction per spinal level treated.ConclusionThis study demonstrated significantly lower occupational doses compared with values reported in the literature. Real-time active personnel dosimeters contributed to a fast optimization and adoption of protective measures throughout the study. Even though our data include both cone-beam computed tomography for navigation planning and intraoperative screw placement verification, we find low patient radiation exposure levels compared with published data.Level Of Evidence3.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.