• Am J Emerg Med · Nov 1994

    Comparative Study Clinical Trial Controlled Clinical Trial

    Prehospital blood pressures: inaccuracies caused by ambulance noise?

    • N H Prasad, L H Brown, S C Ausband, O Cooper-Spruill, R G Carroll, and T W Whitley.
    • Division of Emergency Medical Services, East Carolina University School of Medicine, Greenville, NC 27858.
    • Am J Emerg Med. 1994 Nov 1; 12 (6): 617-20.

    AbstractBlood pressure measurements in a moving ambulance can be difficult to obtain. Sirens, engine noise, and road noise can all interfere with the accurate detection of a patient's blood pressure. This study was undertaken to determine the influence of ambulance noise and vibration on auscultated blood pressures. A model was developed that used dynamic pressures to simulate systolic Korotkoff sounds. Forty-nine emergency personnel were asked to obtain blood pressures using the model in both a quiet environment and in a moving ambulance. A total of 485 blood pressure measurements were obtained. Systolic pressures were randomized to two settings: 76 mm Hg and 138 mm Hg. Stationary readings were compared with moving readings using analysis of variance for repeated measures. Systolic blood pressure measurements obtained in the quiet environment averaged 133 +/- 5 mm Hg at the high setting, and 45 +/- 6 mm Hg at the low setting. Systolic blood pressure measurements obtained in a moving ambulance averaged 86 +/- 7 mm Hg at the high setting, and 41 +/- 7 mm Hg at the low setting. The average differences between quiet and moving measurements were 47 mm Hg at the "high" setting (P < .01) and 4 mm Hg at the "low" setting (P > .01). At physiological levels, blood pressures obtained in moving ambulances differ significantly from those obtained in a quiet environment, which may be caused by road noise and ambulance motion.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…