Dynamic neutralisation of the lumbar spine confirmed on a new lumbar spine simulator in vitro.
- S Freudiger, G Dubois, and M Lorrain.
- snf.ifb@swissonline.ch
- Arch Orthop Trauma Surg. 1999 Jan 1; 119 (3-4): 127-32.
AbstractA new dynamic neutralisation system for lumbar spine segments has been developed and tested on four cadaveric lumbar spines. Segments L4/5 (3 cases) and L3/4 (1 case) were tested on a new lumbar spine simulator which allowed the simultaneous application of bending moments, compressive and shear loads. The average applied loads were 18.3 Nm flexion moment, 2296 N compressive and 458 N anterior shear load for flexion, and 12.5 Nm extension moment, 667 N compressive and 74 N posterior shear load for extension. The relative motion of the upper vertebra in respect to the lower vertebra was measured with the three-dimensional FASTRAK system, using an advanced computer software. The endplate centres as well as the centre of the screw heads were taken as reference points, identified by orthogonally taken radiographs. The dynamic neutralisation system described reduces bending angles and horizontal translations, but it expands vertical translations. The bulging of the posterior annulus is also reduced.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.