    Assessment of spontaneous neuromuscular recovery: A comparison of the TOF-Cuff® with the TOF Watch SX®.

    • Eve Sfeir Machado, Gleicy Keli-Barcelos, Elise Dupuis-Lozeron, Martin R Tramèr, and Christoph Czarnetzki.
    • Division of Anaesthesiology, Department of Anaesthesiology, Pharmacology, Intensive Care and Emergency Medicine, Geneva University Hospitals, Geneva, Switzerland.
    • Acta Anaesthesiol Scand. 2020 Feb 1; 64 (2): 173-179.

    BackgroundTOF-Cuff® is a modified blood pressure cuff used to monitor neuromuscular block. We compared the assessment of spontaneous neuromuscular recovery between TOF-Cuff® (test device) and TOF Watch SX® (reference device).MethodsForty patients aged 18-65 years undergoing elective surgery were enrolled. TOF-Cuff® was installed on an upper arm and the TOF Watch SX® on the thumb of the opposite side. Anaesthesia was induced and maintained with intravenous propofol and sufentanil. After induction, the devices were calibrated and continuous train-of-four (TOF) stimulation was started. A single intravenous dose of rocuronium (0.6 mg kg-1 ) was administered for intubation. The primary outcome was total recovery time (time in minutes from the injection of rocuronium to a normalized TOF ratio of 90%). Agreement between the two devices was calculated using mean difference and limits of agreement.ResultsThe primary outcome could be analysed in 27 patients because of 13 exclusions due to neuromuscular block reversal for shorter procedure surgical time, necessity of reinjection of rocuronium or technical failures of one of the two devices. Median total recovery time with the test device was 45 minutes (interquartile range [IQR] 38.5-61.5) and 63 minutes (IQR 51.1-74.5) with the reference device. Total recovery time with the test device was on average 16.4 minutes shorter (limits of agreement, -6.1 to 39); increasing total recovery time was associated with increasing difference. The TOF ratio of the reference device was on average 0.59 (SD 0.23) when the test device indicated complete recovery. The TOF ratio of the test device was on average 0.98 (SD 0.03) when the reference device indicated complete recovery.ConclusionWhen compared with the TOF Watch SX® , TOF-Cuff® overestimates spontaneous recovery of a rocuronium-induced neuromuscular block.© 2019 The Acta Anaesthesiologica Scandinavica Foundation. Published by John Wiley & Sons Ltd.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…