Eur. J. Intern. Med. · Mar 2007
A comparative study of an ambulatory blood pressure measuring device and a wrist blood pressure monitor with a position sensor versus a mercury sphygmomanometer.
- Sekip Altunkan, Yasemin Genç, and Erkan Altunkan.
- Hypertension Division, Metropol Medical Center, Nisan Sokak, No: 7, 06400, Dikmen, Ankara, Turkey.
- Eur. J. Intern. Med. 2007 Mar 1; 18 (2): 118-23.
BackgroundSelf-measurements of blood pressure (BP) and 24-hour BP measurements are better predictors of cardiovascular mortality and morbidity than office BP measurements. The objective of this study was to compare the accuracy and precision of a wrist BP monitor with a position sensor (Omron 637IT) and of an ambulatory BP measuring monitor (ABPM; Nissei DS-250) with a mercury sphygmomanometer.MethodsA total of 139 patients (69 women and 70 men) were included in the study. The BP of each subject was first measured with a mercury device using the same (left) arm. After this, the wrist monitor was used for BP measurement. Upon completion of the BP readings, 24-hour BP monitoring was performed using Nissei DS-250 monitors. Mean and standard deviations were calculated for all devices. In order to assess the agreement between the measurement methods, the Bland-Altman method and graphics were utilized.ResultsThe mean systolic BP measured by the mercury device was 133.2+/-18.4 mmHg and the diastolic BP was 85.4+/-12.5 mmHg, whereas the digital device measured systolic BP as 135.7+/-17.2 mmHg and diastolic BP as 87.0+/-12.5 mmHg. The 24-hour BP measurement was 134.6+/-16.6 mmHg for systolic BP and 85.6+/-11.1 mmHg for diastolic BP. The difference with regard to systolic BP between the mercury and the Omron devices was -2.5+/-5.3 mmHg, which is within the AAMI standard. However, while the mean values of the differences between the mercury and ABPM devices remained under 5 mmHg, their standard deviation was above +/- 8 mmHg. For diastolic BP, the difference between all of the devices was below 5+/-8 mmHg.ConclusionsThe wrist BP monitor produced results consistent with those of the mercury sphygmomanometer when both were compared with the results of the ABPM. As BP measurement with these devices is a practical and repeatable method, they can be used instead of ABPM in the diagnosis and monitoring of hypertension. However, there is a need for further comparative studies.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.