• Anaesth Crit Care Pain Med · Apr 2019

    Comparative Study Observational Study

    Comparison of Proaqt/Pulsioflex® and oesophageal Doppler for intraoperative haemodynamic monitoring during intermediate-risk abdominal surgery.

    • Grégoire Weil, Cyrus Motamed, Alexandre Eghiaian, Xavier Monnet, and Stéphanie Suria.
    • Gustave-Roussy, Department of anaesthesia, 94805 Villejuif, France.
    • Anaesth Crit Care Pain Med. 2019 Apr 1; 38 (2): 153-159.

    ObjectiveTo compare cardiac index (CI) between Proaqt/PulsioFlex® and oesophageal Doppler (OD) and the ability of the PulsioFlex® to track CI changes induced by fluid challenge and secondly to assess the impact of the time interval between two auto-calibrations of PulsioFlex® on the accuracy of the measured CI.MethodsIn a single hospital, 49 intermediate-risk oncologic abdominal surgery patients were included in an observational study. We measured the cardiac Index (CI) provided by OD and by the Proaqt/PulsioFlex® before and after internal calibration, which were performed randomly at specific intervals after the initial one (30, 60, 90 and 120min). The ability to track fluid responsiveness was evaluated by measuring stroke volume variation, pulse pressure variation (PPV) and CI before and after a 250ml fluid challenge and assessed by a receiver operating characteristic curve analysis.ResultsThe percentages of error before calibration were 51, 58, 82, 81% for 30, 60, 90 and 120min, they were 39, 57, 65, and 54% after calibration. Trending ability is assumed by a 93% concordance rate after applying a 15% exclusion zone. The trend interchangeability rate was 13.75%. The area under the curve for fluid responsiveness measured by PPV and SVV PulsioFlex were respectively 0.67 [0.57-0.77], P<0.01 and 0.75 [0.47-0.66], which was not clinically relevant.ConclusionsThe Proaqt/Pulsioflex® system is not equivalent to OD for haemodynamic monitoring during non-vascular abdominal surgery in intermediate-risk patients. More studies are required to define the effect of the auto-calibration on the system.Copyright © 2018 Société française d'anesthésie et de réanimation (Sfar). Published by Elsevier Masson SAS. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…