Clinical TrialValidation of the Natus CO-Stat End Tidal Breath Analyzer in children and adults.
- H J Vreman, R J Wong, P Harmatz, A A Fanaroff, B Berman, and D K Stevenson.
- Departments of Pediatric, Stanford University School of Medicine, Stanford, CA 94305-5208, USA. henk.vreman@stanford.edu
- J Clin Monit Comput. 1999 Dec 1; 15 (7-8): 421-7.
ObjectiveThe performance of a point-of-care, noninvasive end tidal breath carbon monoxide analyzer (CO-Stat End Tidal Breath Analyzer, Natus Medical Inc.) that also reports end tidal carbon dioxide (ETCO2) and respiratory rate (RR), was compared to established, marketed (predicate) devices in children (n = 39) and adults (n = 48) who are normal or at-risk of elevated CO excretion.MethodsConcentrations of end tidal breath CO (ETCO), room air CO, ETCO corrected for inhaled CO (ETCOc), ETCO2, and RR were measured with the CO-Stat analyzer and the data compared to those obtained from the same subjects using the Vitalograph BreathCO monitor (Vitalograph, Inc.) for ETCOc and the Pryon CO2 monitor (SC210 and SC300, Pryon Corp) for ETCO2 and RR. Adults and children were studied at three medical centers. The data were analyzed by paired t-tests and linear regression. Bias and imprecision between the CO-Stat analyzer and the predicate devices was calculated by the method of Bland and Altman.ResultsPaired t-tests, performed on the three parameters measured with the CO-Stat analyzer and predicate devices showed that only the ETCOc values in the adults and the ETCO2 values in the children were significantly different (lower, p < or = 0.0001, and higher, p < or = 0.0001, respectively). The mean bias and imprecision of the CO-Stat analyzer for adult ETCOc and children ETCO2 measurements were -0.9 +/- 1.2 ppm and 0.4 +/- 0.6%, respectively. Linear regression analysis for the ETCOc results in children and adults had a high degree of correlation (r = 0.91 and 0.98, respectively).ConclusionsWe conclude that in a clinical environment the Natus CO-Stat End Tidal Breath Analyzer performs at least as well as predicate devices for the measurements of ETCOc, ETCO2, and RR.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.