Response time of the Opti-Q continuous cardiac output pulmonary artery catheter in the urgent mode to a step change in cardiac output.
- L J Goldstein.
- Department of Surgery, University of California, Davis-East Bay, Oakland, CA 94602, USA. goldstein@pol.net
- J Clin Monit Comput. 1999 Dec 1; 15 (7-8): 435-9.
ObjectivesThis study was conducted to determine the response time of the Opti-Q continuous cardiac output (CCO) device to a step change in cardiac.DesignProspective study.SettingUniversity hospital animal lab.ModelFemale sheep.InterventionsIn ten animals, cardiac output was altered suddenly by opening and closing a peripheral arteriovenous shunt to test the response time of the CCO system.Measurements And Main ResultsCardiac output was measured continuously by thermodilution and ultrasonic techniques while an arteriovenous shunt was opened and closed. A total of 53 dynamic observations were made (5-6 per animal). The mean response time of the continuous cardiac output device was 86 seconds and was unaffected by the magnitude or direction of the change in cardiac output. It was also unaffected by the animal's weight. CCO values were not statistically different from standard thermodilution measurement (p = 0.895). Shunt flow ranged from 430 to 1730 ml/min and averaged 812 ml/min. The mean CCO with the shunt closed was 4.62 L/min. There was 1.5 to 2 minutes under or overshoot in cardiac output in 11% of the measurements.ConclusionsContinuous cardiac output measurement was as accurate as those made by standard bolus thermodilution. The average response time to acute changes in cardiac output was approximately 1.5 minutes or ten times faster than previously reported systems. Response time is independent of animal mass, shunt volume and the direction of cardiac output perturbations.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.