• Anesthesia and analgesia · Aug 2020

    Comparative Study

    Comparison of Hemoglobin Measurements by 3 Point-of-Care Devices With Standard Laboratory Values and Reliability Regarding Decisions for Blood Transfusion.

    • Marianne Johnson, Peter C Marwick, and Johan F Coetzee.
    • From the Department of Anesthesiology and Critical Care, Faculty of Medicine and Health Sciences, Stellenbosch University and Tygerberg Academic Hospital, Cape Town, South Africa.
    • Anesth. Analg. 2020 Aug 1; 131 (2): 640-649.

    BackgroundWe compared the accuracy of 3 point-of-care testing (POCT) devices with central laboratory measurements and the extent to which between-method disagreements could influence decisions to transfuse blood.MethodsHemoglobin concentrations [Hb] were measured in 58 adult patients undergoing cardiothoracic surgery using 2 Ilex GEM Premier 3500 blood gas analyzers (BG_A and BG_B) and a HemoCue Hb-201 device (HemoCue). Measurements were compared with our central laboratory's Siemens Advia 2120 flow cytometry system (laboratory [Hb] [Lab[Hb]]), regarded as the gold standard. We considered that between-method [Hb] differences exceeding 10% in the [Hb] range 6-10 g/dL would likely erroneously influence erythrocyte transfusion decisions.ResultsThe 70 Lab[Hb] measurements ranged from 5.8 to 16.7 g/dL, of which 25 (36%) were <10.0 g/dL. Measurements by all 4 devices numbered 57. Mean POCT measurements did not differ significantly (P > .99). Results of the Bland-Altman analyses revealed statistically significant bias, with predominant underestimations by all 3 POCTs predominating. HemoCue upper and lower limits of agreement (LOA) were narrower, and the 95% confidence intervals (95% CIs) of the LOAs did not overlap with those of BG_A and BG_B. Similarly, a narrow mountain plot demonstrated greater precision for the HemoCue. Comparing BG_A with BG_B revealed no bias and narrow LOA. Error grid analysis within the [Hb] range 6-10 g/dL revealed that 5.3% of HemoCue measurements were beyond the permissible 10.0% error zone in contrast to 19.0% and 16.0% of the blood gas measurements. Possible inappropriate transfusion decisions based on POCT values generally erred toward unnecessary transfusions. Calculations of Cohen κ statistic indicated better chance-corrected agreement between HemoCue and Lab[Hb] regarding erythrocyte transfusions than the blood gas analyzers.ConclusionsAll 3 POCT devices underestimated the Lab[Hb] and cannot be used interchangeably with standard laboratory measurements. BG_A and BG_B can be considered to be acceptably interchangeable with each other. Whereas the HemoCue had little bias and good precision, the blood gas analyzers revealed large bias and poor precision. We conclude that the tested HemoCue provides more reliable measurements, especially within the critical 6-10 g/dL range, with reduced potential for transfusion errors. Decisions regarding erythrocyte transfusions should also be considered in the light of clinical findings.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.