    Comparison of arterial blood pressure measurements obtained invasively or oscillometrically using a Datex S/5 Compact monitor in anaesthetised adult horses.

    • Tyfane T Yamaoka, Derek Flaherty, Patricia Pawson, Marian Scott, and Adam Auckburally.
    • Institute of Biodiversity, Animal Health and Comparative Medicine, School of Veterinary Medicine, University of Glasgow, Glasgow, UK. Electronic address: Tyfane.Yamaoka@glasgow.ac.uk.
    • Vet Anaesth Analg. 2017 May 1; 44 (3): 492-501.

    ObjectiveTo assess agreement between noninvasive blood pressure (NIBP) oscillometrically-derived values from a multiparameter monitor (Datex Ohmeda S/5 Compact) with those obtained by invasive blood pressure (IBP) measurement in anaesthetised horses undergoing elective surgery.Study DesignProspective clinical study.AnimalsA total of 40 healthy adult horses.MethodsHorses were anaesthetised with various anaesthetic protocols (based on clinical requirements). Depending on positioning, cannulation of the facial or lateral metatarsal artery was performed for IBP measurement. The cannula was connected via a transducer to the monitor. An appropriately sized NIBP cuff was placed around the tail base and connected to the same monitor. Systolic (SAP), mean (MAP) and diastolic (DAP) arterial blood pressures were continuously recorded from the invasive system, and at 3 minute intervals from the oscillometric system, throughout the surgical procedure using a Datex iCollect program. An appropriate arithmetic correction factor was applied to the oscillometric results where the cuff was not level with the heart. Assessment of the degree of agreement between invasive and noninvasive readings at each time point was performed using a modified Bland-Altman analysis.ResultsWhile in many horses there was relatively close correlation between the values obtained over time, there was substantial variability in individual animals which resulted in wide Bland-Altman limits of agreement. The oscillometric device over-reads by approximately 32, 23 and 22 mmHg, and under-reads by 26, 17 and 19 mmHg for SAP, MAP and DAP, respectively, compared with the IBP values. However, using the mean difference and standard deviation, the device conforms to American College of Veterinary Internal Medicine (ACVIM) standards.Conclusions And Clinical RelevanceOscillometric blood pressure measurement using the Datex Ohmeda S/5 Compact multiparameter monitor conforms to ACVIM standards when the NIBP cuff is placed on the tail. However, because of the wide variability in measurements, we cannot recommend this technique to guide therapy in anaesthetised adult horses.Copyright © 2017 Association of Veterinary Anaesthetists and American College of Veterinary Anesthesia and Analgesia. Published by Elsevier Ltd. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.