• Neurocritical care · Aug 2020

    Comparative Study Observational Study

    Comparison of Three Point-of-Care Ultrasound Views and MRI Measurements for Optic Nerve Sheath Diameter: A Prospective Validity Study.

    • Ronak Raval, Jay Shen, Deon Lau, Nick Ferguson, Thomas Kelly, Justin Daniels, Ihab Dorotta, and Davinder Ramsingh.
    • Department of Anesthesiology, Loma Linda University Medical Center, Loma Linda, CA, USA.
    • Neurocrit Care. 2020 Aug 1; 33 (1): 173-181.

    IntroductionPoint-of-care ultrasound of the optic nerve sheath diameter (ONSD) to diagnose increased intracranial pressure (ICP) is of great interest in various clinical scenarios. Yet, the lack of examination standardization has made clinical utility difficult. We compare three ultrasound ocular plane views (inferior, sagittal, and transverse), which are currently used in the literature to evaluate their consistency. Comparisons for each view to magnetic resonance imaging (MRI) measurements were also made.MethodsFifty-one patients with recent MRI of the brain, but without clinical or radiological signs of elevated ICP, were selected to undergo ocular sonography via three ultrasound planes (inferior, sagittal, and transverse). Optic nerve sheath was measured in each ultrasound view as well with MRI. Image quality scores were assigned for the ultrasound views in different orientations. The three ocular plane views were analyzed for correlation. In addition, correlation of the three ocular ultrasound views with MRI was also performed.ResultsCorrelation analysis showed a wide variability in the correlation between different ultrasound views with magnitude range of 0.1 to 0.8 and directions being both positive and negative. There was a difference in image quality scores between the ultrasound views. The inferior and transverse orientations were superior to the sagittal orientation in achieving high image quality. Comparison to MRI measurements did not demonstrate a significant correlation.ConclusionOur findings suggest that absolute measurements should not be compared across different ultrasound orientations given the wide variability in the correlation between the ultrasound views used to assess the optic nerve sheath. The inferior and transverse ultrasound views are the most likely to yield high-quality images, although the specific view, for the best image, in an individual patient can vary. We would caution against absolute values of ONSD to indicate increased ICP, as it may be view dependent.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.