• Vet Anaesth Analg · Jul 2014

    Comparative Study

    Comparison of an infrared anaesthetic agent analyser (Datex-Ohmeda) with refractometry for measurement of isoflurane, sevoflurane and desflurane concentrations.

    • Andrea S Rudolff, Yves P S Moens, Bernd Driessen, and Tamas D Ambrisko.
    • Anaesthesiology and perioperative Intensive-Care Medicine, Department for Companion Animals and Horses, University of Veterinary Medicine, Vienna, Austria.
    • Vet Anaesth Analg. 2014 Jul 1; 41 (4): 386-92.

    ObjectiveTo assess agreement between infrared (IR) analysers and a refractometer for measurements of isoflurane, sevoflurane and desflurane concentrations and to demonstrate the effect of customized calibration of IR analysers.Study DesignIn vitro experiment.SubjectsSix IR anaesthetic monitors (Datex-Ohmeda) and a single portable refractometer (Riken).MethodsBoth devices were calibrated following the manufacturer's recommendations. Gas samples were collected at common gas outlets of anaesthesia machines. A range of agent concentrations was produced by stepwise changes in dial settings: isoflurane (0-5% in 0.5% increments), sevoflurane (0-8% in 1% increments), or desflurane (0-18% in 2% increments). Oxygen flow was 2 L minute(-1) . The orders of testing IR analysers, agents and dial settings were randomized. Duplicate measurements were performed at each setting. The entire procedure was repeated 24 hours later. Bland-Altman analysis was performed. Measurements on day-1 were used to yield calibration equations (IR measurements as dependent and refractometry measurements as independent variables), which were used to modify the IR measurements on day-2.ResultsBias ± limits of agreement for isoflurane, sevoflurane and desflurane were 0.2 ± 0.3, 0.1 ± 0.4 and 0.7 ± 0.9 volume%, respectively. There were significant linear relationships between differences and means for all agents. The IR analysers became less accurate at higher gas concentrations. After customized calibration, the bias became almost zero and the limits of agreement became narrower.Conclusions And Clinical RelevanceIf similar IR analysers are used in research studies, they need to be calibrated against a reference method using the agent in question at multiple calibration points overlapping the range of interest.© 2013 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesia and Analgesia.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…