ObjectivesTo compare the biomechanical stability of internal fixation of extra-articular, dorsally unstable distal radius fractures fixed by 1 of 2 methods, either a standard dorsal nonlocked T-plate or a volar locked fixed-angle plate.DesignBiomechanical cadaveric study.SettingBiomechanical testing laboratory.InterventionIn 6 matched pairs of fresh-frozen cadaveric specimens, a simulated unstable extra-articular distal radius fracture was created. The fractures were stabilized with either a dorsal 3.5-mm stainless steel T-plate or a titanium locked volar fixed-angle plate. Specimens were axially loaded at 5 points (centrally, volarly, dorsally, radially, and ulnarly) and then cyclically loaded for 5000 cycles with an 80 N central load. Postcyclical loading, specimens were once again axially loaded at the 5 points.Main Outcome MeasuresInitial fixation stiffness and stiffness after midaxial cyclical loading was compared at the 5 points.ResultsWith the volar locked fixed-angle plate, fixation was significantly stiffer than with the dorsal nonlocked T-plate for ulnar and volar loading in single-cycle testing. After cyclic loading, the locked volar fixed-angle plate maintained more of its initial stiffness than the dorsal nonlocked T-plate. The dorsal 3.5-mm stainless steel T-plate's stiffness when dorsally loaded significantly decreased after cyclical loading.ConclusionsThe volar locked fixed-angle plate maintained a greater percentage of its initial stiffness after cyclic loading compared to the dorsal nonlocked plate. Also, the volar locked plate was stiffer than the dorsal nonlocked plate for all loading configurations tested except when subjected to a dorsally applied eccentric load.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.