    Airway Occlusion Pressure as an Estimate of Respiratory Drive and Inspiratory Effort During Assisted Ventilation.

    • Irene Telias, Detajin Junhasavasdikul, Nuttapol Rittayamai, Lise Piquilloud, Lu Chen, Niall D Ferguson, Ewan C Goligher, and Laurent Brochard.
    • Interdepartmental Division of Critical Care Medicine and.
    • Am. J. Respir. Crit. Care Med. 2020 May 1; 201 (9): 1086-1098.

    AbstractRationale: Monitoring and controlling respiratory drive and effort may help to minimize lung and diaphragm injury. Airway occlusion pressure (P0.1) is a noninvasive measure of respiratory drive.Objectives: To determine 1) the validity of "ventilator" P0.1 (P0.1vent) displayed on the screen as a measure of drive, 2) the ability of P0.1 to detect potentially injurious levels of effort, and 3) how P0.1vent displayed by different ventilators compares to a "reference" P0.1 (P0.1ref) measured from airway pressure recording during an occlusion.Methods: Analysis of three studies in patients, one in healthy subjects, under assisted ventilation, and a bench study with six ventilators. P0.1vent was validated against measures of drive (electrical activity of the diaphragm and muscular pressure over time) and P0.1ref. Performance of P0.1ref and P0.1vent to detect predefined potentially injurious effort was tested using derivation and validation datasets using esophageal pressure-time product as the reference standard.Measurements and Main Results: P0.1vent correlated well with measures of drive and with the esophageal pressure-time product (within-subjects R2 = 0.8). P0.1ref >3.5 cm H2O was 80% sensitive and 77% specific for detecting high effort (≥200 cm H2O ⋅ s ⋅ min-1); P0.1ref ≤1.0 cm H2O was 100% sensitive and 92% specific for low effort (≤50 cm H2O ⋅ s ⋅ min-1). The area under the receiver operating characteristics curve for P0.1vent to detect potentially high and low effort were 0.81 and 0.92, respectively. Bench experiments showed a low mean bias for P0.1vent compared with P0.1ref for most ventilators but precision varied; in patients, precision was lower. Ventilators estimating P0.1vent without occlusions could underestimate P0.1ref.Conclusions: P0.1 is a reliable bedside tool to assess respiratory drive and detect potentially injurious inspiratory effort.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.